Robust Magnetic Field Homogeneity Enhancement via Local Passive Shimming in a Superconducting Magnet for Animal MRI
- 22 April 2026
- journal article
- research article
- Published by Wiley in NMR in Biomedicine
- Vol. 39 (6) , e70298
Abstract
In ultrahigh‐field magnetic resonance imaging (MRI) systems, achieving subppm static magnetic field homogeneity is critical. However, engineering tolerances and assembly errors often introduce complex field inhomogeneities that global shimming alone cannot always fully correct. In this study, a novel local passive shimming method is developed for small‐bore MRI superconducting magnets. By establishing a linear programming model for local shimming, the approach enables region‐specific correction of high‐order field variations. Numerical simulations and experimental verification on a 7 T/300‐mm superconducting magnet demonstrate significant improvements in homogeneity, reducing root mean square error (RMSE) and peak‐to‐peak variation across the diameter of spherical volume (DSV). The optimized shim configuration achieved a peak‐to‐peak inhomogeneity of 4.23 ppm within the DSV, indicating substantial enhancement over baseline performance. These results indicate that the proposed technique serves as a practical and efficient complement to global shimming strategies, offering a robust solution for field correction in small‐bore MRI systems.Keywords
Funding Information
- National Natural Science Foundation of China (52277031)
- National Natural Science Foundation of China (52293423)
This publication has 26 references indexed in Scilit:
- Automatic Magnetic Field Shimming by Spectrum Convergence Method: Evaluation at a 7‐T Animal MRI SystemNMR in Biomedicine, 2025
- An L0‐Norm–Based Method for Passive Shimming Design in MRI: A Simulation‐Based StudyNMR in Biomedicine, 2025
- Active shimming for a 25 T NMR superconducting magnet by spectrum convergence methodJournal of Magnetic Resonance, 2024
- A novel passive shimming scheme using explicit control of magnetic field qualities with minimal use of ferromagnetic materialsMagnetic Resonance in Medicine, 2022
- A novel passive shimming method for the correction of magnetic fields above the patient bed in MRIJournal of Magnetic Resonance, 2015
- Magnetic field homogenization of the human prefrontal cortex with a set of localized electrical coilsMagnetic Resonance in Medicine, 2009
- Combined passive and active shimming for in vivo MR spectroscopy at high magnetic fieldsJournal of Magnetic Resonance, 2006
- Sample-specific diamagnetic and paramagnetic passive shimmingJournal of Magnetic Resonance, 2006
- Mitigation of susceptibility‐induced signal loss in neuroimaging using localized shim coilsMagnetic Resonance in Medicine, 2005
- Optimization of static field homogeneity in human brain using diamagnetic passive shimsMagnetic Resonance in Medicine, 2002