A miniaturized catheter 2-D array for real-time, 3-D intracardiac echocardiography
Authors:
- W. LeeWLW. Lee
- S.F. IdrissSIS.F. Idriss
- P.D. WolfPWP.D. Wolf
- S.W. SmithSSS.W. Smith
- 1 November 2004
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 51 (10) , 1334-1346
Abstract
The design, fabrication, and characterization of a 112 channel, 5 MHz, two-dimensional (2-D) array transducer constructed on a six layer flexible polyimide interconnect circuit is described. The transducer was mounted in a 7 Fr (2.33 mm outside diameter) catheter for use in real-time intracardiac volumetric imaging. Two transducers were constructed: one with a single silver epoxy matching layer and the other without a matching layer. The center frequency and -6 dB fractional bandwidth of the transducer with a matching layer were 4.9 MHz and 31%, respectively. The 50 /spl Omega/ pitch-catch insertion loss was 80 dB, and the typical interelement crosstalk was -30 dB. The final element yield was greater than 97% for both transducers. The transducers were used to acquire real-time, 3-D images in an in vivo sheep model. We present in vivo images of cardiac anatomy obtained from within the coronary sinus, including the left and right atria, aorta, coronary arteries, and pulmonary veins. We also present images showing the manipulation of a separate electrophysiological catheter into the coronary sinus.Keywords
This publication has 30 references indexed in Scilit:
- A minimally invasive ultrasound probe using non-coax cabling2002
- Quantitative assessment of left ventricular perfusion defects using real-time three-dimensional myocardial contrast echocardiographyJournal of the American Society of Echocardiography, 2002
- Intracardiac Echocardiography: Newest TechnologyJournal of the American Society of Echocardiography, 2000
- Real-time three-dimensional echocardiography for measurement of left ventricular volumesThe American Journal of Cardiology, 1999
- Intracardiac doppler hemodynamics and flow: new vector, phased-array ultrasound-tipped catheterThe American Journal of Cardiology, 1999
- Two-dimensional arrays for medical ultrasound using multilayer flexible circuit interconnectionIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1998
- High-speed ultrasound volumetric imaging system. II. Parallel processing and image displayIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1991
- High-speed ultrasound volumetric imaging system. I. Transducer design and beam steeringIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1991
- Acoustic properties of particle/polymer composites for ultrasonic transducer backing applicationsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1990
- A contrast‐detail analysis of diagnostic ultrasound imagingMedical Physics, 1982