A-scan ultrasound in ophthalmology: A simulation tool
- 21 September 2021
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 97 (1) , 18-24
Abstract
In the present study, we developed a computational tool for simulating the ophthalmological applications of A-scan ultrasound, including cataract characterisation and biometry. A-scan biometry is used to measure the axial length (AL) of the eye before cataract surgery to calculate the refractive power of the intraocular lens to be implanted. Errors in the measurement of the AL lead to post-surgical refractive errors. The simulation tool was developed using the k-Wave Matlab toolbox, together with a user-friendly interface developed in Matlab. Diverse error sources were evaluated. Constant ultrasound speed assumptions may introduce refractive errors of up to 0.6 D; by contrast, probe positioning errors had a lower impact, of up to 0.11 D. The correct identification of the Bruch's membrane is limited not only by axial resolution constraints but also by the low reflection coefficient at the retina/choroid interface. Regarding cataract characterisation, the amplitudes of the echoes reflected at the lens interfaces are sensitive to diverse cataract types and severities, and a more realistic representation could be obtained by using a higher resolution in the eye grid; however, the required computational times would make simulations impracticable when using personal computers. The simulation tool shows good versatility for evaluating diverse aspects of A-scan biometry.Keywords
Funding Information
- European Regional Development Fund (PTDC/SAU-SER/28758/2017)
- Fundação para a Ciência e a Tecnologia (PTDC/SAU-SER/28758/2017)
This publication has 30 references indexed in Scilit:
- A Review of Tissue Substitutes for Ultrasound ImagingUltrasound in Medicine & Biology, 2010
- k-Wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fieldsJournal of Biomedical Optics, 2010
- Density of Ocular Components of the Bovine EyeOptometry and Vision Science, 2009
- Measurements of attenuation coefficient for evaluating the hardness of a cataract lens by a high-frequency ultrasonic needle transducerPhysics in Medicine & Biology, 2009
- ABSORPTION OF ULTRASOUND IN THE AQUEOUS HUMOUR, LENS AND VITREOUS BODYActa Ophthalmologica, 2009
- Biometry and intraocular lens power calculationCurrent Opinion in Ophthalmology, 2008
- Calculation of intraocular lens power: a reviewActa Ophthalmologica Scandinavica, 2007
- Acoustic velocity and attenuation of eye tissues at 20 MHzUltrasound in Medicine & Biology, 1994
- Acoustic parameters of ocular tissuesUltrasound in Medicine & Biology, 1985
- MEASUREMENT OF THE VELOCITY OF SOUND IN SOME PARTS OF THE EYEActa Ophthalmologica, 1958