Design, construction and traceable calibration of a phosphor-based fibre-optic thermometer from 0 °C to 650 °C
Open Access
- 12 March 2021
- journal article
- research article
- Published by IOP Publishing in Measurement Science and Technology
- Vol. 32 (9) , 094004
Abstract
Measurement traceability, in which there is a documented chain of calibration links from a measurement back to a primary realisation of a unit, is vital for achieving the smallest uncertainty in metrology. A sensor that can provide a traceable link in harsh environments may be advantageous even if its uncertainty is larger than that of typical calibrated devices. A phosphor-based thermometer that has optical rather than electrical connections has been built and then calibrated in terms of the International Temperature Scale of 1990. Its standard uncertainty ranges from 0.5 °C at 0 °C to 2 °C at 650 °C. While this uncertainty cannot match what is possible with thermocouples or resistance thermometers, the optical basis of the sensor means it can be used in situations where electrical connections are precluded.Keywords
Funding Information
- Horizon 2020 Framework Programme (17IND04)
This publication has 10 references indexed in Scilit:
- Optimization of synthesis and compositional parameters of magnesium germanate and fluoro-germanate thermographic phosphorsJournal of Alloys and Compounds, 2018
- Temperature measurement techniques for gas and liquid flows using thermographic phosphor tracer particlesProgress in Energy and Combustion Science, 2018
- On surface temperature measurements with thermographic phosphors: A reviewProgress in Energy and Combustion Science, 2013
- Thermographic phosphors for thermometry: A survey of combustion applicationsProgress in Energy and Combustion Science, 2011
- Thermographic Phosphors for High Temperature Measurements: Principles, Current State of the Art and Recent ApplicationsSensors, 2008
- Review of luminescent based fibre optic temperature sensorsSensor Review, 2005
- Remote thermometry with thermographic phosphors: Instrumentation and applicationsReview of Scientific Instruments, 1997
- Development of a high-temperature fiber-optic thermometer probe using fluorescent decayReview of Scientific Instruments, 1991
- The International Temperature Scale of 1990 (ITS-90)Metrologia, 1990
- Kinetics of Crystallization of Stoichiometric SiO 2 Glass in H 2 O AtmospheresJournal of the American Ceramic Society, 1966