Improvement of Hemocompatibility in Centrifugal Blood Pump With Hydrodynamic Bearings and Semi‐open Impeller: In Vitro Evaluation
- 20 October 2009
- journal article
- Published by Wiley in Artificial Organs
- Vol. 33 (10) , 798-804
Abstract
We have developed a noncontact‐type centrifugal blood pump with hydrodynamic bearings and a semi‐open impeller for mechanical circulatory assist. The impeller is levitated by an original spiral‐groove thrust bearing and a herringbone‐groove journal bearing, without any additional displacement‐sensing module or additional complex control circuits. The pump was improved by optimizing the groove direction of the spiral‐groove thrust bearing and the pull‐up magnetic force between the rotor magnet and the stator coil against the impeller. To evaluate hemocompatibility, we conducted a levitation performance test and in vitro hemocompatibility tests by means of a mock‐up circulation loop. In the hemolysis test, the normalized index of hemolysis was reduced from 0.721 to 0.0335 g/100 L corresponding to an expansion of the bearing gap from 1.1 to 56.1 µm. In the in vitro antithrombogenic test, blood pumps with a wide thrust bearing gap were effective in preventing thrombus formation. Through in vitro evaluation tests, we confirmed that hemocompatibility was improved by balancing the hydrodynamic fluid dynamics and magnetic forces.Keywords
This publication has 11 references indexed in Scilit:
- Antithrombogenic Properties of a Monopivot Magnetic Suspension Centrifugal Pump for Circulatory AssistArtificial Organs, 2008
- Hemocompatibility of a hydrodynamic levitation centrifugal blood pumpJournal of Artificial Organs, 2007
- Hemolytic Performance of a MagLev Disposable Rotary Blood Pump (MedTech Dispo): Effects of MagLev Gap Clearance and Surface RoughnessArtificial Organs, 2006
- Third‐generation Blood Pumps With Mechanical Noncontact Magnetic BearingsArtificial Organs, 2006
- SIMPLE IN VITRO TESTING METHOD FOR ANTITHROMBOGENIC EVALUATION OF CENTRIFUGAL BLOOD PUMPSASAIO Journal, 2006
- The Pivot Wash in Two Impeller Modes for the Baylor/Miwatec Centrifugal Blood PumpArtificial Organs, 2005
- Current Status of the Gyro Centrifugal Blood Pump—Development of the Permanently Implantable Centrifugal Blood Pump as a Biventricular Assist Device (NEDO Project)Artificial Organs, 2004
- Testing of a Centrifugal Blood Pump With a High Efficiency Hybrid Magnetic BearingASAIO Journal, 2003
- The HeartMate III: Design and In Vivo Studies of a Maglev Centrifugal Left Ventricular Assist DeviceArtificial Organs, 2001
- MAGNETIC SUSPENSION CONTROLS FOR CONTINUOUS FLOW VENTRICULAR ASSIST DEVICE CFVAD IIIASAIO Journal, 1997