Repeatability and Agreement of Two Swept-Source Optical Coherence Tomographers for Anterior Segment Parameter Measurements
Authors:
- 20 January 2022
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Glaucoma
- Vol. 31 (7) , 602-608
Abstract
Purpose: The aim was to evaluate the repeatability and agreement of two swept-source optical coherence tomographers for anterior segment parameters in healthy subjects. Patients and Methods: Automated mark of scleral spur and angle recess for the CASIA2 and manual mark of scleral spur and angle recess for the Anterion were performed, and then the measurement values of the related parameters were automatically generated. Subjects with pupil diameter (PD) differing by <15% between the 2 devices were included. PD, lens vault (LV), anterior chamber depth (ACD), angle to angle distance (ATA), anterior chamber width (ACW), anterior chamber angle (ACA), angle opening distance (AOD), and trabecular iris space area (TISA) were measured in the horizontal images with both CASIA2 and Anterion. Intraclass correlation coefficient (ICC) was used to evaluate intradevice repeatability. Bland-Altman plots were performed to assess the agreement between the 2 devices. Results: Thirty-five right eyes of 35 subjects were included with a mean age of 25.60±3.00 years. The CASIA2 showed moderate to good intradevice repeatability (ICCs ranged from 0.786 to 0.989) whereas the Anterion showed good intradevice repeatability (ICCs ranged from 0.921 to 0.998) for anterior segment parameters. Compared with the CASIA2, the Anterion offered larger values of ACA500/750, AOD500/750, and TISA500/750 but smaller values of PD, LV, ACD, ATA, and ACW. Good agreement for PD, ACD, ATA, and ACW was detected with 95% limits of agreement of −1.02 to 1.02 mm, −0.11 to 0.14 mm, −0.17 to 0.19 mm, −0.13 to 0.28 mm, respectively. Poor agreement for LV, ACA500/750, AOD500/750, and TISA500/750 was achieved with the 2 devices. Conclusions: Anterion outperformed CASIA2 on intradevice repeatability. While agreement was noted for some parameters using manual Anterion and automated CASIA2 approaches, poor agreement of LV and angle parameters indicates that measurements from these optical coherence tomography devices should not be considered interchangeable.Keywords
This publication has 33 references indexed in Scilit:
- Swept source optical coherence tomography measurement of the iris–trabecular contact (ITC) index: a new parameter for angle closureGraefe's Archive for Clinical and Experimental Ophthalmology, 2012
- Swept source / Fourier domain polarization sensitive optical coherence tomography with a passive polarization delay unitOptics Express, 2012
- Dynamic Analysis of Iris Configuration with Anterior Segment Optical Coherence TomographyInvestigative Ophthalmology & Visual Science, 2010
- Anterior segment biometry with 2 imaging technologies: Very-high-frequency ultrasound scanning versus optical coherence tomographyJournal of Cataract & Refractive Surgery, 2008
- Dynamic Analysis of Dark–Light Changes of the Anterior Chamber Angle with Anterior Segment OCTInvestigative Ophthalmology & Visual Science, 2007
- Direct Measurement of the Ciliary Sulcus Diameter by 35-Megahertz Ultrasound BiomicroscopyOphthalmology, 2007
- Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomographyBritish Journal of Ophthalmology, 2007
- Defining "occludable" angles in population surveys: drainage angle width, peripheral anterior synechiae, and glaucomatous optic neuropathy in east Asian peopleBritish Journal of Ophthalmology, 2004
- Forming inferences about some intraclass correlation coefficients.Psychological Methods, 1996
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986