Diagnostic Accuracy of Pulmonary CT Angiography at Low Tube Voltage: Intraindividual Comparison of a Normal-Dose Protocol at 120 kVp and a Low-Dose Protocol at 80 kVp Using Reduced Amount of Contrast Medium in a Simulation Study

Authors:

Abstract
OBJECTIVE. The purpose of this study was to simulate pulmonary emboli (PE) and image quality at low tube energy and reduced contrast material volume in normal-dose pulmonary CT angiography (CTA) images and to analyze the diagnostic accuracy with normal- and low-dose pulmonary CTA. MATERIALS AND METHODS. Normal-dose pulmonary CTA examinations using 120 kVp and 100 mL of contrast material in 10 patients with no PE were retrospectively selected. The image characteristics of an 80-kVp low-dose pulmonary CTA protocol (patient exposure reduction, 57%) with 75 mL of contrast material were simulated. Four different sets of filling defects were computer simulated in identical locations in each normal-dose and corresponding low-dose examination, equaling 783 PE in 40 normal-dose and 40 low-dose datasets. Ten normal-dose and 10 low-dose examinations contained no emboli and were used as controls. The 100 pulmonary CTA studies were randomly assessed by three readers blinded to PE location and image quality. The result...