In vivo safe catheter visualization and slice tracking using an optically detunable resonant marker
Open Access
- 23 September 2004
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 52 (4) , 860-868
Abstract
The purpose of this study was to test the in vivo feasibility of safe automatic catheter tracking based on an optically detunable resonant marker installed on the catheter tip, and also to test the compatibility of this approach with guidewire materials. The design of the resonant marker and the integration into the real‐time MR environment is described. The catheter was used for real‐time MR‐guided catheterization of the aorta, left ventricle, and carotid in two swine. For in‐plane visualization, the marker was repeatedly detuned. For automatic slice tracking, a projection difference measurement including detuning was interleaved with the imaging sequence. In vitro experiments were conducted to investigate the RF‐safety of the marker and the effect of the guidewires on the signal intensity. For all orientations the marker provided excellent in vivo contrast using a radial steady‐state free‐precession sequence. Flashing of the marker by repetitive tuning/detuning further improved the in‐plane visualization. Automatic slice tracking during real‐time imaging was successfully performed. The plastic guidewires did not interfere with the marker, and detuning by guidewires containing nitinol could be compensated. In conclusion, automatic slice tracking as well as excellent in‐plane visualization can be achieved with this approach and it is safe with respect to RF transmission. Magn Reson Med 52:860–868, 2004.Keywords
This publication has 19 references indexed in Scilit:
- Safety of MRI‐guided endovascular guidewire applicationsJournal of Magnetic Resonance Imaging, 2000
- Heating Around Intravascular Guidewires by Resonating RF WavesJournal of Magnetic Resonance Imaging, 2000
- Reduction of resonant RF heating in intravascular catheters using coaxial chokesMagnetic Resonance in Medicine, 2000
- Catheter visualization using locally induced, actively controlled field inhomogeneitiesMagnetic Resonance in Medicine, 1997
- Guidewire antennas for MR fluoroscopyMagnetic Resonance in Medicine, 1997
- Intravascular magnetic resonance imaging using a loopless catheter antennaMagnetic Resonance in Medicine, 1997
- Towards active guidewire visualization in interventional magnetic resonance imagingMagnetic Resonance Materials in Physics, Biology and Medicine, 1996
- Vascular interventions guided by ultrafast MR imaging: Evaluation of different materialsMagnetic Resonance in Medicine, 1994
- Real‐time position monitoring of invasive devices using magnetic resonanceMagnetic Resonance in Medicine, 1993
- Magnetic Susceptibility Effects and Their Application in the Development of New Ferromagnetic Catheters for Magnetic Resonance ImagingInvestigative Radiology, 1990