Comparison of dosimetric verification results and plan complexity for treatment plans generated in Ethos 1.0 and Ethos 2.0 systems
Authors:
- Paulina WesolowskaPWPaulina Wesolowska
- Joanna Iwanowska-HankeJIJoanna Iwanowska-Hanke
- Anna PaciorkiewiczAPAnna Paciorkiewicz
- Sandra MaluszczakSMSandra Maluszczak
- Mikołaj TarchalskiMTMikołaj Tarchalski
- Bartłomiej SadowskiBSBartłomiej Sadowski
- Marta FillmannMFMarta Fillmann
- Agnieszka WalewskaAWAgnieszka Walewska
Open Access
- 18 November 2026
- journal article
- Published by Medical Journals Sweden AB in Acta Oncologica
- Vol. 65, 888-893
Abstract
Introduction: The Varian Ethos adaptive radiotherapy system was upgraded from version 1.0 to 2.0, introducing an enhanced multi-leaf collimator (MLC) dose calculation model and a new optimization engine. We evaluated whether this affected agreement between calculated and measured dose, and plan complexity. Material and methods: Twenty-four plans (14 intensity-modulated radiotherapy [IMRT], 10 volumetric modulated arc therapy [VMAT]; 26 measurement configurations) covering 10 anatomical sites were analyzed in three matched configurations: original Ethos 1.0 plans (E1.0), the same plans recalculated in Ethos 2.0 with the new MLC model only (E1.0→2.0), and plans re-optimized in Ethos 2.0 (E2.0). Quality assurance comprised ionization chamber point-dose measurement and global gamma analysis on an ArcCHECK array at 3%/3 mm, 3%/2 mm and 2%/2 mm (10% threshold). Complexity was quantified with UCoMX v1.1; groups were compared with the Wilcoxon signed-rank test.Results: Mean signed chamber deviation was −1.59 ± 0.54% (E1.0), −1.54 ± 0.57% (E1.0→2.0) and −1.39 ± 0.63% (E2.0), with no significant differences (p > 0.6); all met the ±3% criterion. Gamma pass rates at 3%/3 mm were 99.2 ± 0.9%, 98.7 ± 1.0% (p < 0.001) and 98.6 ± 2.1% (p = 0.059). Re-optimized plans used 12% fewer monitor units (1,169 ± 395 vs. 1,338 ± 471, p < 0.001) and were simpler (MCSv 0.31 to 0.35; SAS10 mm 0.27 to 0.20; both p < 0.001).Conclusion: The upgrade preserved quality-assurance compliance and produced simpler, lower-monitor-unit plans, but the systematic −1.5% offset between calculated and measured point dose persisted, showing the new MLC model alone does not remove it.This publication has 15 references indexed in Scilit:
- Dosimetric evaluation of Ethos 2.0 high‐fidelity mode for single‐isocenter SRSJournal of Applied Clinical Medical Physics, 2025
- Improving efficiency in lung SAbR planning using integrated tools for X‐ray based adaptive radiotherapyJournal of Applied Clinical Medical Physics, 2025
- Commissioning of Halcyon enhanced leaf model in the Eclipse treatment planning system: Focus on simple slit fields and VMAT dose calculationJournal of Applied Clinical Medical Physics, 2025
- Evaluating machine learning enhanced intelligent‐optimization‐engine (IOE) performance for ethos head‐and‐neck (HN) plan generationJournal of Applied Clinical Medical Physics, 2023
- Do we need patient‐specific QA for adaptively generated plans? Retrospective evaluation of delivered online adaptive treatment plans on Varian EthosJournal of Applied Clinical Medical Physics, 2022
- Assessment of efficacy in automated plan generation for Varian Ethos intelligent optimization engineJournal of Applied Clinical Medical Physics, 2022
- Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithmJournal of Applied Clinical Medical Physics, 2022
- On the reduction of aperture complexity in kidney SABRJournal of Applied Clinical Medical Physics, 2021
- Experience in commissioning the halcyon linacMedical Physics, 2019
- Characterization of the HalcyonTM multileaf collimator systemJournal of Applied Clinical Medical Physics, 2019