Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm
Authors:
Open Access
- 9 January 2022
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 23 (3) , e13519
Abstract
Purpose: The dual‐layer multi‐leaf collimator (MLC) in Halcyon involves further complexities in the dose calculation process, because the leaf‐tip transmission varies according to the leaf trailing pattern. For the volumetric modulated arc therapy (VMAT) treatment, the prescribed dose for the target volume can be sensitive to the leaf‐tip transmission change. This report evaluates the dosimetric consequence due to the uncertainty of the dual‐layer MLC model in Eclipse through the dose verifications for clinical VMAT. Additionally, the Halcyon leaf‐tip model is empirically adjusted for the VMAT dose calculation with the Acuros XB.Materials and methods: For this evaluation, an in‐house program that analyzes the leaf position in each layer was developed. Thirty‐two clinical VMAT plans were edited into three leaf sequences: dual layer (original), proximal single layer, or distal single layer. All leaf sequences were verified using Delta4 according to the dose difference (DD) and the global gamma index (GI). To improve the VMAT dose calculation accuracy, the dosimetric leaf gap (DLG) was adjusted to minimize the DD in single‐layer leaf sequences.Results: The mean of DD were −1.35%, −1.20%, and −1.34% in the dual‐layer, proximal single‐layer, and distal single‐layer leaf sequences, respectively. The changes in the mean of DD between leaf sequences were within 0.2%. However, the calculated doses differed from the measured doses by approximately 1% in all leaf sequences. The tuned DLG was increased by 0.8 mm from the original DLG in Eclipse. When the tuned DLG was used in the dose calculation, the mean of DD neared 0% and GI with a criterion of 2%/2 mm yielded a pass rate of more than 98%.Conclusion: No significant change was confirmed in the dose calculation accuracy between the leaf sequences. Therefore, it is suggested that the dosimetric consequence due to the leaf trailing was negligibly small in clinical VMAT plans. The DLG tuning for Halcyon can be useful for reducing the dose calculation uncertainties in Eclipse VMAT and required in the commissioning for Acuros XB.Keywords
Funding Information
- Japan Society for the Promotion of Science (JP19K17186)
This publication has 17 references indexed in Scilit:
- Single-arc VMAT optimization for dual-layer MLCPhysics in Medicine & Biology, 2019
- Varian Halcyon Dosimetric Comparison for Multi-Arc VMAT Prostate and Head-and-Neck CancersMedical Dosimetry, 2019
- Density scaling of phantom materials for a 3D dose verification systemJournal of Applied Clinical Medical Physics, 2018
- A Practical Method to Optimize Quality Assurance Results of Arc Therapy Plans in Beam ModelingJournal of Medical Physics, 2018
- Tolerance limits and methodologies forIMRTmeasurement‐based verificationQA:Recommendations ofAAPMTask Group No. 218Medical Physics, 2018
- Optimization of the dosimetric leaf gap for use in planning VMAT treatments of spine SABR casesJournal of Applied Clinical Medical Physics, 2017
- Intrafractional Baseline Shift or Drift of Lung Tumor Motion During Gated Radiation Therapy With a Real-Time Tumor-Tracking SystemInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2016
- On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plansJournal of Applied Clinical Medical Physics, 2014
- Verification of dosimetric accuracy on the TrueBeam STx: Rounded leaf effect of the high definition MLCMedical Physics, 2012
- Observations on Real-Time Prostate Gland Motion Using Electromagnetic TrackingInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008