Failure Patterns After Linked Semiconstrained Total Elbow Arthroplasty for Posttraumatic Arthritis
Authors:
- Peter ZarkadasPZPeter Zarkadas
- Joaquin Sanchez-SoteloJSJoaquin Sanchez-Sotelo
- Bernard MorreyBMBernard Morrey
- 1 June 2010
- journal article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 92 (6) , 1432-1441
Abstract
Ve semiconstrained total elbow arthroplasties for the treatment of posttraumatic arthritis. Sixty-nine elbows with a retained primary prosthesis were followed for an average of nine years. Clinical results were graded with use of the Mayo Elbow Performance Score. Radiographs were assessed for mechanical failure, and all complications were recorded. Results: Sixteen primary arthroplasties (19%) failed. Causes of failure included isolated bushing wear (seven), infection (four), component fracture (three), and component loosening (two). The most common cause of early failure (failure after less than five years) was infection, whereas intermediate-term failure (failure after five to ten years) typically was due to bushing wear. Late failure (failure after more than ten years) was uncommon and involved component loosening or fracture. Seventy-five percent of the failures were in patients who were less than sixty years old at the time of surgery (p = 0.03). Progressive radiolucent lines were noted around four implants, three of which had clinically important loosening. Total elbow arthroplasty was associated with significant improvements in terms of pain, motion, and the Mayo Elbow Performance Scores (p ≤ 0.002). Sixty-eight percent of the patients achieved a good or excellent clinical result, and 74% were subjectively satisfied. Kaplan-Meier analysis demonstrated a fifteen-year survival rate of 70% with revision or resection for any reason as the end point. Conclusions: Semiconstrained total elbow arthroplasty in patients with posttraumatic arthritis places high demands on the implant and is associated with a relatively high failure rate. Seventy-five percent of failures occur in patients less than sixty years of age, and infection continues to represent a frequent mode of early failure. Bushing wear and component loosening or fracture are seen more commonly in the intermediate and late term, whereas aseptic loosening remains relatively uncommon. Level of Evidence: Therapeutic Level IV. See Instructions to Authors for a complete description of levels of evidence....Keywords
This publication has 20 references indexed in Scilit:
- Modes of Wear After Semiconstrained Total Elbow ArthroplastyJournal of Bone and Joint Surgery, 2008
- Coonrad-Morrey total elbow replacement for primary and revision surgery: A 2- to 7.5-year follow-up studyJournal of Shoulder and Elbow Surgery, 2007
- Fracture and nonunion of the olecranon in total elbow arthroplastyJournal of Shoulder and Elbow Surgery, 2006
- Long-term results with the Kudo type 3 total elbow arthroplastyJournal of Shoulder and Elbow Surgery, 2006
- Total elbow arthroplasty with the Coonrad/Coonrad-Morrey prosthesisThe Journal of Bone and Joint Surgery. British volume, 2006
- Outcomes of Total Elbow Arthroplasty for Rheumatoid Arthritis: Comparative Study of Three ImplantsJournal of Bone and Joint Surgery, 2005
- Polyethylene Wear After Total Elbow ArthroplastyJournal of Bone and Joint Surgery, 2005
- Total elbow arthroplasty failure due to overuse, C-ring failure, and/or bushing wearJournal of Shoulder and Elbow Surgery, 2005
- Limited influence of prosthetic position on aseptic loosening of elbow replacementsActa Orthopaedica, 2005
- Complex fractures of the distal humerus in the elderlyThe Journal of Bone and Joint Surgery. British volume, 2002