Improved precision with Hologic Apex software
Open Access
- 29 March 2008
- journal article
- research article
- Published by Springer Nature in Osteoporosis International
- Vol. 19 (11) , 1597-1602
Abstract
Summary The precision of Hologic Apex v2.0 analysis software is significantly improved from Hologic Delphi v11.2 software and is comparable to GE Lunar Prodigy v7.5 software. Apex and Delphi precisions were, respectively, 1.0% vs. 1.2% (L1-L4 spine), 1.l % vs. 1.3% (total femur), 1.6% vs. 1.9% (femoral neck), and 0.7% vs. 0.9% (dual total femur). Introduction Precision of bone mineral density (BMD) measurements by dual-energy X-ray absorptiometry (DXA) is known to vary by manufacturer, model, and technologist. This study evaluated the precision of three analysis versions: Apex v2.0 and Delphi v11.2 (Hologic, Inc.), and Prodigy v7.5 (GE Healthcare, Inc.) independent of technologist skill. Methods Duplicate spine and dual hip scans on 90 women were acquired on both Delphi and Prodigy DXA systems at three clinics. BMD measures were converted to standardized BMD (sBMD) units. Precision errors were described as a root-mean-square (RMS) standard deviations and RMS percent coefficients of variation across the population. Results Apex and Delphi values were highly correlated (r ranged from 0.90 to 0.99). Excluding the right neck, the Apex precision error was found to be 20% to 25% lower than the Delphi (spine: 1.0% versus 1.2% (p < 0.05), total hip: 1.1% versus 1.3% (p < 0.05), right neck: 2.3% versus 2.6% (p > 0.1)). No statistically significant differences were found in the precision error of the Apex and Prodigy (p > 0.05) except for the right neck (2.3% versus 1.8% respectively, p = 0.03). Conclusion The Apex software has significantly lower precision error compared to Delphi software with similar mean values, and similar precision to that of the Prodigy.Keywords
This publication has 9 references indexed in Scilit:
- A study of the long-term precision of dual-energy X-ray absorptiometry bone densitometers and implications for the validity of the least-significant-change calculationOsteoporosis International, 2006
- Comparison of BMD precision for Prodigy and Delphi spine and femur scansOsteoporosis International, 2006
- Cross-Calibration and Minimum Precision Standards for Dual-Energy X-ray Absorptiometry: The 2005 ISCD Official PositionsJournal of Clinical Densitometry, 2006
- Standardization of Bone Mineral Density at Femoral Neck, Trochanter and Ward's TriangleOsteoporosis International, 2001
- Monitoring Skeletal Changes by Radiological TechniquesJournal of Bone and Mineral Research, 1999
- Universal Standardization of Bone Density Measurements: A Method with Optimal Properties for Calibration Among Several InstrumentsJournal of Bone and Mineral Research, 1997
- Accurate assessment of precision errors: How to measure the reproducibility of bone densitometry techniquesOsteoporosis International, 1995
- Rates of change in bone mineral density of the spine, heel, femoral neck and radius in healthy postmenopausal womenBone and Mineral, 1992
- The clinical application of serial bone mass measurementsBone and Mineral, 1991