An efficient technique for estimating the two-dimensional temperature distributions around multiple cryo-surgical probes based on combining contributions of unit circles
Authors:
- 11 February 2016
- journal article
- research article
- Published by Taylor & Francis in Computer Methods in Biomechanics and Biomedical Engineering
- Vol. 19 (13) , 1462-1474
Abstract
This study presents an efficient, fast and accurate method for estimating the two-dimensional temperature distributions around multiple cryo-surgical probes. The identical probes are inserted into the same depth and are operated simultaneously and uniformly. The first step in this method involves numerical derivation of the temporal performance data of a single probe, embedded in a semi-infinite, tissue-like medium. The results of this derivation are approximated by algebraic expressions that form the basis for computing the temperature distributions of multiple embedded probes by combining the data of a single probe. Comparison of isothermal contours derived by this method to those computed numerically for a variety of geometrical cases, up to 15 inserted probes and 2–10 min times of operation, yielded excellent results. Since this technique obviates the solution of the differential equations of multiple probes, the computational time required for a particular case is several orders of magnitude shorter than that needed for obtaining the full numerical solution. Blood perfusion and metabolic heat generation rates are demonstrated to inhibit the advancement of isothermal fronts. Application of this method will significantly shorten computational times without compromising the accuracy of the results. It may also facilitate expeditious consideration of the advantages of different modes of operation and the number of inserted probes at the early design stage.Keywords
This publication has 36 references indexed in Scilit:
- Computerized planning of prostate cryosurgery using variable cryoprobe insertion depthCryobiology, 2010
- Computerized planning of cryosurgery using bubble packing: An experimental validation on a phantom materialInternational Journal of Heat and Mass Transfer, 2008
- Cryosurgery planning using bubble packing in 3DComputer Methods in Biomechanics and Biomedical Engineering, 2008
- Cryosurgery for TumorsJournal of the American College of Surgeons, 2007
- An efficient numerical technique for bioheat simulations and its application to computerized cryosurgery planningComputer Methods and Programs in Biomedicine, 2007
- Retroperitoneoscopy-Assisted Cryoablation of Renal Tumors Using Multiple 1.5mm Ultrathin Cryoprobes: A Preliminary ReportEuropean Urology, 2005
- Cryosurgery in aggressive, benign, and low-grade malignant bone tumoursThe Lancet Oncology, 2005
- Cryotherapeutic ablation of liver tumoursBritish Journal of Surgery, 2002
- CryosurgeryAnnual Review of Biomedical Engineering, 2000
- Thermal Aspects of CryosurgeryJournal of Heat Transfer, 1976