Development of a Biomimetic Extensor Mechanism for Restoring Normal Kinematics of Finger Movements Post-Stroke
- 30 August 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Vol. 27 (10) , 2107-2117
Abstract
Cable-driven devices for hands allow compact and lightweight design that could provide various functional movements. However, for many patients post-stroke, cable-driven devices produce nonphysiologic movements, such as metacarpophalangeal joint hyperextension, due to their abnormal passive joint impedance. In this study, we developed a novel bio-inspired device mimicking the anatomy of the extensor mechanism of the human finger, which can be tuned for individuals to provide `subject-specific' assistance to achieve physiological movement patterns. We first evaluated the proposed design via mathematical modeling and computer simulation. Its performance was then tested experimentally with twenty-four subjects, including six healthy and eighteen chronic stroke survivors. We determined the loading condition of the device from the experimental identification of passive joint impedance of each subject before device use. Our results showed that the proposed design could achieve improved spatiotemporal coordination of finger movements compared to conventional cable-driven design by providing `subject-specific' assistance based on identified passive stiffness values of each subject. We also identified a significant (negative) correlation between the metacarpophalangeal joint stiffness and the intrinsic exotendon loading level across subjects. The proposed system can restore normal movement patterns for patients with different types of impairments, which were previously found important in improving rehabilitative outcomes.Keywords
Funding Information
- National Research Foundation of Korea
- National Research Foundation of Korea (NRF-2016R1A5A1938472, NRF-2017R1A2B4011961)
- Convergence Technology Development Program for Bionic Arm through the National Research Foundation of Korea
- National Research Foundation of Korea (NRF-2015M3C1B2052817)
- National Science Foundation (CBET-1452763)
This publication has 38 references indexed in Scilit:
- Robot-assisted rehabilitation of hand functionCurrent Opinion in Neurology, 2010
- Development and pilot testing of HEXORR: Hand EXOskeleton Rehabilitation RobotJournal of NeuroEngineering and Rehabilitation, 2010
- Spasticity Mechanisms – for the ClinicianFrontiers in Neurology, 2010
- What Do Motor “Recovery” and “Compensation” Mean in Patients Following Stroke?Neurorehabilitation and Neural Repair, 2008
- Effect of Finger Posture on the Tendon Force Distribution Within the Finger Extensor MechanismJournal of Biomechanical Engineering, 2008
- Robot-based hand motor therapy after strokeBrain, 2007
- Central slip tenotomy for the treatment of chronic mallet finger: an anatomic studyThe Journal of Hand Surgery, 2004
- Chronic stress tolerance levels for human articular cartilage: Two nonuniform contact models applied to long-term follow-up of CDHJournal of Biomechanics, 1995
- Architectural design of the human intrinsic hand musclesThe Journal of Hand Surgery, 1992
- A dynamic model for finger interphalangeal coordinationJournal of Biomechanics, 1988