Comparison of Astigmatism Prediction Accuracy for Toric Lens Implantation from Two Swept-Source Optical Coherence Tomography Devices
Open Access
- 1 November 2022
- journal article
- research article
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. ume 16, 3795-3802
Abstract
Purpose: To compare the astigmatism prediction accuracy for toric intraocular lens (IOL) implantation between two swept-source optical coherence tomography (SS-OCT) devices: Argos (Movu, a Santec Company) and the IOLMaster 700 (Carl Zeiss Meditec). Methods: This was a retrospective chart review of 59 eyes (44 patients) with corneal astigmatism and cataract that underwent cataract surgery or refractive lens exchange surgery with a toric IOL. Biometry was performed on all patients prior to cataract surgery and the Barrett toric IOL calculator was used. Visual acuity was measured postoperatively. Manifest refraction was measured at 1 month and compared to the predicted postoperative residual refraction. Preoperative K measurements between devices were also compared. Results: Mean cylinder prediction error was − 0.17 ± 0.43 for Argos and 0.12 ± 0.56 for IOLMaster 700. The cylinder prediction error 0.5 D or less was not significantly different between the devices, with 83.1% (49 eyes) for Argos and 76.3% (45 eyes) for IOLMaster 700 (p = 0.206). Spherical equivalent prediction error was 0.13 ± 0.39 for Argos and 0.25 ± 0.50 for IOLMaster 700. The mean spherical equivalent prediction error 0.5 D or less was significantly different between the devices, with 79.7% (47 eyes) for Argos and 61.0% (36 eyes) for IOLMaster 700 (p = 0.016). Conclusion: The prediction accuracies were similar between the devices, except for spherical equivalent, where a higher percentage of eyes were 0.5 D or less of the prediction with the Argos compared to the IOLMaster.Keywords
This publication has 17 references indexed in Scilit:
- Measurement of Posterior Corneal Astigmatism by the IOLMaster 700Journal of Refractive Surgery, 2018
- Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometryPLOS ONE, 2018
- Multicenter study of optical low-coherence interferometry and partial-coherence interferometry optical biometers with patients from the United States and ChinaJournal of Cataract & Refractive Surgery, 2016
- Biometry measurements using a new large-coherence–length swept-source optical coherence tomographerJournal of Cataract & Refractive Surgery, 2016
- Clinical comparison of a new swept-source optical coherence tomography–based optical biometer and a time-domain optical coherence tomography–based optical biometerJournal of Cataract & Refractive Surgery, 2015
- Prevalence of corneal astigmatism before cataract surgery in Chinese patientsJournal of Cataract & Refractive Surgery, 2013
- Visual performance in cataract patients with low levels of postoperative astigmatism: full correction versus spherical equivalent correctionClinical Ophthalmology, 2012
- Prevalence of corneal astigmatism in patients having routine cataract surgery at a teaching hospital in the United KingdomJournal of Cataract & Refractive Surgery, 2011
- Comparison of 2 laser instruments for measuring axial lengthJournal of Cataract & Refractive Surgery, 2010
- Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliabilityJournal of Cataract & Refractive Surgery, 2001