Multicenter study of optical low-coherence interferometry and partial-coherence interferometry optical biometers with patients from the United States and China
- 1 January 2016
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 42 (1) , 62-67
Abstract
To evaluate the agreement between the measurements provided by a new optical biometer, the Aladdin, based on optical low-coherence interferometry (OLCI), and those provided by the most commonly used optical biometer (IOLMaster 500), based on partial-coherence interferometry (PCI). Multicenter clinical trial. Prospective evaluation of diagnostic test. In this study, 2 samples of adult patients were enrolled, 1 in the United States and the other in China. The U.S. group included a sample of consecutive patients scheduled for cataract surgery. The China group included a sample of healthy subjects with no cataracts. In both cases, only 1 eye of each patient was analyzed. Axial length (AL), corneal power (in diopters [D]) (K), anterior chamber depth (ACD) (corneal epithelium to lens), and corneal astigmatism were measured. All values were analyzed using a paired t test, the Pearson product-moment correlation coefficient (r), and Bland-Altman plots. In the U.S. and China groups, the OLCI mean AL values did not show a statistically significant difference from PCI values and showed excellent agreement and correlation. On the contrary, OLCI measured a lower mean K (-0.14 D) and a deeper ACD measurements (U.S. +0.16 mm and China +0.05 mm). These differences were statistically significant (P < .0001). Vector analysis did not show a statistically significant difference in astigmatism measurements. Agreement between OLCI and PCI was good. However, the small but statistically significant differences in K and ACD measurements make constant optimization necessary when calculating the intraocular lens power using theoretical formulas. Dr. Hoffer licenses the registered trademark name Hoffer to Carl Zeiss-Meditec (PCI), Haag-Streit (Lenstar), Movu (Argos), Oculus (Pentacam, AXL), Nidek (AL-Scan), Tomey (OA-2000), Topcon EU Visia Imaging (Aladdin), Ziemer (Galilei G6), and all A-scan biometer manufacturers. Dr. Shammas licenses his formulas to Carl Zeiss-Meditec (PCI), Haag-Streit (Lenstar), Nidek (AL-Scan), and Topcon EU (Visia Imaging) (Aladdin). None of the other authors has a financial or proprietary interest in any material or method mentioned.Keywords
This publication has 11 references indexed in Scilit:
- Evaluation of a new optical biometry device for measurements of ocular components and its comparison with IOLMasterBritish Journal of Ophthalmology, 2014
- Validity and repeatability of the Aladdin ocular biometerBritish Journal of Ophthalmology, 2013
- Comparison of 2 laser instruments for measuring axial lengthJournal of Cataract & Refractive Surgery, 2010
- Comparison of Corneal Powers Obtained from 4 Different DevicesAmerican Journal of Ophthalmology, 2009
- A new optical low coherence reflectometry device for ocular biometry in cataract patientsBritish Journal of Ophthalmology, 2009
- Relationship between anterior corneal asphericity and refractive variablesGraefe's Archive for Clinical and Experimental Ophthalmology, 2008
- Immersion A-scan compared with partial coherence interferometryJournal of Cataract & Refractive Surgery, 2002
- Power vector analysis of the optical outcome of refractive surgeryJournal of Cataract & Refractive Surgery, 2001
- Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to HaigisGraefe's Archive for Clinical and Experimental Ophthalmology, 2000
- A three-part system for refining intraocular lens power calculationsJournal of Cataract & Refractive Surgery, 1988