Biomechanical Evaluation of a Double-Threaded Pedicle Screw in Elderly Vertebrae

Authors:
  • SHSean M. Haddock
    Sean M. Haddock
  • MLMichael A. K. Liebschner
    Michael A. K. Liebschner
  • TKTony M. Keaveny
    Tony M. Keaveny
  • WRWilliam S. Rosenberg
    William S. Rosenberg

Abstract
We sought to test the hypothesis that a pedicle screw that has two parallel threads of different heights throughout the full length of the screw could increase both bone purchase and pull-out strength compared with a standard single-threaded screw of similar dimensions. A single-threaded pedicle screw and a double-threaded pedicle screw were respectively placed into the paired pedicles of 21 vertebral bodies. The screws were then pulled out of the pedicles, and output parameters were measured. Although insertional torque was, on average, 14.5% higher (p = 0.039) for the single-threaded screw, maximum pull-out strength (p = 0.12), energy-to-failure (p = 0.39), and stiffness (p = 0.54) were not statistically different for the two screw types. It is concluded that a second, smaller inner thread on a double-threaded pedicle screw does not translate into either increased bone purchase or higher pull-out strengths.

This publication has 25 references indexed in Scilit: