Similarly low blood metal ion levels at 10‐years follow‐up of total hip arthroplasties with Oxinium, CoCrMo, and stainless steel femoral heads. Data from a randomized clinical trial
Open Access
- 10 November 2022
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research Part B: Applied Biomaterials
- Vol. 111 (4) , 821-828
Abstract
The use of inert head materials such as ceramic heads has been proposed as a method of reducing wear and corrosion products from the articulating surfaces in total hip arthroplasty, as well as from the stem‐head taper connection. The aim of the present study was to compare the blood metal ion levels in patients with Oxinium and CoCrMo modular femoral heads, as well as monoblock stainless steel Charnley prostheses at 10 years postoperatively. The 150 patients with osteoarthritis of the hip joint included in a randomized clinical trial were grouped according to femoral head material. One group (n = 30) had received the Charnley monoblock stainless steel stem (DePuy, UK). The other patients (n = 120) received a Spectron EF CoCrMo stem with either a 28 mm CoCrMo or Oxinium modular head (Smith & Nephew, USA). After 10 years, 38 patients had withdrawn, 19 deceased, 7 revised due to aseptic loosening and 5 revised due to infection. The 81 patients with median age of 79 years (70–91) were available for whole blood metal ion analysis. The levels of Co, Cr, Ni and Zr in the blood were generally low with all the head materials (medians p = .2–.8). Based on the low blood metal ion values in our study groups, no indication of severe trunnion corrosion in patients with CoCrMo heads was observed, neither was there any beneficial reduction in metal ion exposure with the Oxinium femoral heads.Keywords
Funding Information
- Helse Vest
This publication has 29 references indexed in Scilit:
- A Systematic Review of Systemic Cobaltism After Wear or Corrosion of Chrome-Cobalt Hip ImplantsJournal of Patient Safety, 2019
- Trunnionosis in total hip arthroplasty: a reviewJournal of Orthopaedics and Traumatology, 2016
- Metal Ion Concentrations in Young, Active Patients Following Total Hip Arthroplasty with the Use of Modern Bearing CouplesThe Journal of Arthroplasty, 2015
- Oxinium femoral head damage generated by a metallic foreign body within the polyethylene cup following recurrent dislocation episodesOrthopaedics & Traumatology: Surgery & Research, 2013
- Do Ceramic Femoral Heads Reduce Taper Fretting Corrosion in Hip Arthroplasty? A Retrieval StudyClinical Orthopaedics and Related Research, 2013
- The Influence of Head Size on Corrosion and Fretting Behaviour at the Head-Neck Interface of Artificial Hip JointsThe Journal of Arthroplasty, 2013
- Coordinating Retrieval and Register Studies Improves Postmarket Surveillance2012
- Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip ArthroplastyJournal of Bone and Joint Surgery, 2012
- A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral components using radiostereometric analysis at 2 yearsActa Orthopaedica, 2011
- Wear and migration of highly cross‐linked and conventional cemented polyethylene cups with cobalt chrome or Oxinium femoral heads: A randomized radiostereometric study of 150 patientsJournal of Orthopaedic Research, 2011