Wear and migration of highly cross‐linked and conventional cemented polyethylene cups with cobalt chrome or Oxinium femoral heads: A randomized radiostereometric study of 150 patients
Authors:
- 25 February 2011
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 29 (8) , 1222-1229
Abstract
This randomized study was performed to compare wear and migration of five different cemented total hip joint articulations in 150 patients. The patients received either a Charnley femoral stem with a 22.2 mm head or a Spectron EF femoral stem with a 28 mm head. The Charnley articulated with a γ‐sterilized Charnley Ogee acetabular cup. The Spectron EF was used with either EtO‐sterilized non‐cross‐linked polyethylene (Reflection All‐Poly) or highly cross‐linked (Reflection All‐Poly XLPE) cups, combined with either cobalt chrome (CoCr) or Oxinium femoral heads. The patients were followed with repeated RSA measurements for 2 years. After 2 years, the EtO‐sterilized non‐cross‐linked Reflection All‐Poly cups had more than four times higher proximal penetration than its highly cross‐linked counterpart. Use of Oxinium femoral heads did not affect penetration at 2 years compared to heads made of CoCr. Further follow‐up is needed to evaluate the benefits, if any, of Oxinium femoral heads in the clinical setting. The Charnley Ogee was not outperformed by the more recently introduced implants in our study. We conclude that this prostheses still represents a standard against which new implants can be measured. © 2011 Orthopaedic Research Society Published by Wiley Periodicals, Inc. J Orthop Res 29: 1222–1229, 2011Keywords
This publication has 28 references indexed in Scilit:
- The creep and wear of highly cross-linked polyethyleneThe Journal of Bone and Joint Surgery. British volume, 2008
- Effect of femoral head size on polyethylene wear and synovitis after total hip arthroplasty: A sonographic and radiographic study of 39 patientsActa Orthopaedica, 2008
- Surface damage to an Oxinium femoral head prosthesis after dislocationThe Journal of Bone and Joint Surgery. British volume, 2007
- Very low wear of non-remelted highly cross-linked polyethylene cups: An RSA study lasting up to 6 yearsActa Orthopaedica, 2007
- Polyethylene wear rate and osteolysis: Critical threshold versus continuous dose-response relationshipJournal of Orthopaedic Research, 2005
- Guidelines for standardization of radiostereometry (RSA) of implantsActa Orthopaedica, 2005
- Conquest of a Worldwide Human DiseaseClinical Orthopaedics and Related Research, 2004
- Bioactive cement or ceramic/porous coating vs. Conventional cement to obtain early stability of the acetabular cup Randomised study of 96 hips followed with radiostereometryJournal of Orthopaedic Research, 2004
- Documentation of hip prostheses used in NorwayA critical review of the literature from 1996–2000Acta Orthopaedica, 2004
- Accurate accuracy assessment: Review of basic principles (Acta Orthop Scand 1999; 70 (4): 319-321)Acta Orthopaedica, 2000