Harmonisation of quantitative assessment between different generation of PET/CT: Biograph mCT and Biograph Vision
Open Access
- 20 April 2024
- journal article
- research article
- Published by SPRI of Radiation Hygiene Prof. PV Ramzaev in Radiatsionnaya Gygiena = Radiation Hygiene
- Vol. 17 (1) , 44-53
Abstract
The usage of modern positron emission tomography scanners, in particular with digital detectors, allows obtaining images with better quality, increases the detection of small pathological lesions, reduces scanning time and the activity administered to the patient which leads to a decrease of patient dose as well. However, the values of the quantitative image parameters shift upward, which can lead to significant differences with the quantitative assessment obtained on the previous generation device. In order to compare quantitative assessments obtained on different generations of PET/CT, it is necessary to harmonise quantitative image parameters and perform regular quality control. The aim of current work is the comparison of different methods for harmonization of quantitative image parameters on the example of harmonisation of two PET/CT: Biograph mCT 128 and Biograph Vision 600. NEMA IEC Body phantom filled with 18F solution was scanned in Listmode in two bed positions with overlap in the sphere area during five minutes per bed position. Recovery coefficient used for harmonisation was measured for each sphere of the phantom. Harmonisation between Vision and mCT was performed using two methods: choosing of harmonised reconstruction parameters and EQ.PET technology. The acceptable divergence range between the recovery coefficients for Vision and for mCT is ±10% (20% range). The recovery coefficients measured for reconstruction: 4 iterations and 5 subsets, ToF+PSF, Gaussian 7 mm, matrix 220x220 completely fit within the 20% range. The recovery coefficients measured using EQ = 6 mm (optimal value) fit within the 20% range except the spheres with a diameter of 10 and 13 mm. Both harmonisation methods allow to approximate the quantitative assessment/ However, EQ.PET has limitations for the small lesions. Choosing harmonised reconstruction parameters is the mostwidely used harmonisation method; the EQ.PET allows to harmonise quantitative assessment without the use of multiple reconstruction protocols and losses in visualization abilityKeywords
This publication has 15 references indexed in Scilit:
- Pitfalls on PET/CT Due to Artifacts and InstrumentationSeminars in Nuclear Medicine, 2021
- Digital PET/CT allows for shorter acquisition protocols or reduced radiopharmaceutical dose in [18F]-FDG PET/CTAnnals of Nuclear Medicine, 2021
- Phantom-based image quality assessment of clinical 18F-FDG protocols in digital PET/CT and comparison to conventional PMT-based PET/CTEJNMMI Physics, 2020
- How to harmonize SUVs obtained by hybrid PET/CT scanners with and without point spread function correctionPhysics in Medicine & Biology, 2018
- SUV Harmonization Between Different Hybrid PET/CT SystemsClinical Nuclear Medicine, 2018
- Impact of PET/CT system, reconstruction protocol, data analysis method, and repositioning on PET/CT precision: An experimental evaluation using an oncology and brain phantomMedical Physics, 2017
- Generating harmonized SUV within the EANM EARL accreditation program: software approach versus EARL-compliant reconstructionAnnals of Nuclear Medicine, 2016
- FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0European Journal of Nuclear Medicine and Molecular Imaging, 2014
- Improvement in PET/CT image quality in overweight patients with PSF and TOFAnnals of Nuclear Medicine, 2014
- Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patientsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2013