A Monte Carlo based analytic model of the in-room neutron ambient dose equivalent for a Mevion gantry-mounted passively scattered proton system
Authors:
- Milad Baradaran-GhahfarokhiMBMilad Baradaran-Ghahfarokhi
- Francisco ReynosoFRFrancisco Reynoso
- Arash DarafshehADArash Darafsheh
- Baozhou SunBSBaozhou Sun
- Michael T PrusatorMPMichael T Prusator
- Sasa MuticSMSasa Mutic
- Tianyu ZhaoTZTianyu Zhao
- 22 September 2020
- journal article
- research article
- Published by IOP Publishing in Journal of Radiological Protection
- Vol. 40 (4) , 980-996
Abstract
The goal of this study was to develop a Monte Carlo (MC)-based analytical model that can predict the in-room ambient dose equivalent from a Mevion gantry-mounted passively scattered proton system. The Mevion S250 and treatment vault were simulated using the MCNPX MC code. The results of the in-room neutron dose measurements, using an FHT 762 WENDI-II detector, were employed to benchmark the MC-derived values. After tuning the MCNPX MC code, for the same beam delivery parameters, the code was used to calculate the neutron spectra and ambient dose equivalent in the vault and at varying angles from the isocenter. Then, based on the calculations, an analytical model was reconstructed and data were fitted to derive the model parameters at 95% confidence intervals (CI). The MCNPX codes were tuned to within about 19% of the measured values for most of the measurements in the vault. For the maze, up to 0.08 mSv Gy −1 discrepancies were found between the experimental measurements and MCNPX calculated results. The analytical model showed up to 18% discrepancy for distances between 100 and 600 cm from the isocenter compared to the MC calculations. The model may underestimate the neutron ambient dose equivalent up to 21% for distances less than 100 cm from the isocenter. The proposed analytical model can be used to estimate the contribution of the secondary neutron dose from the Mevion S250 for the design of local shielding inside the proton therapy treatment vault.Keywords
This publication has 24 references indexed in Scilit:
- Evaluation of neutron dose equivalent from the Mevion S250 proton accelerator: measurements and calculationsPhysics in Medicine & Biology, 2013
- Analytical model for out-of-field dose in photon craniospinal irradiationPhysics in Medicine & Biology, 2013
- Uncertainties and correction methods when modeling passive scattering proton therapy treatment heads with Monte CarloPhysics in Medicine & Biology, 2011
- An analytic model of neutron ambient dose equivalent and equivalent dose for proton radiotherapyPhysics in Medicine & Biology, 2010
- Neutron equivalent doses and associated lifetime cancer incidence risks for head & neck and spinal proton therapyPhysics in Medicine & Biology, 2009
- Equivalent dose and effective dose from stray radiation during passively scattered proton radiotherapy for prostate cancerPhysics in Medicine & Biology, 2008
- Monte Carlo study of neutron dose equivalent during passive scattering proton therapyPhysics in Medicine & Biology, 2007
- Calculations of neutron dose equivalent exposures from range-modulated proton therapy beamsPhysics in Medicine & Biology, 2005
- Accurate Monte Carlo simulations for nozzle design, commissioning and quality assurance for a proton radiation therapy facilityMedical Physics, 2004
- WENDIHealth Physics, 2000