Biomechanical evaluation of a biomimetic spinal construct
Open Access
- 26 June 2014
- journal article
- research article
- Published by Wiley in Journal of Experimental Orthopaedics
- Vol. 1 (1) , 3
Abstract
Laboratory spinal biomechanical tests using human cadaveric or animal spines have limitations in terms of disease transmission, high sample variability, decay and fatigue during extended testing protocols. Therefore, a synthetic biomimetic spine model may be an acceptable substitute. The goal of current study is to evaluate the properties of a synthetic biomimetic spine model; also to assess the mechanical performance of lateral plating following lateral interbody fusion. Three L3/4 synthetic spinal motion segments were examined using a validated pure moment testing system. Moments (±7.5 Nm) were applied in flexion-extension (FE), lateral bending (LB) and axial rotation (AR) at 1Hz for total 10000 cycles in MTS Bionix. An additional test was performed 12 hours after 10000 cycles. A ±10 Nm cycle was also performed to allow provide comparison to the literature. For implantation evaluation, each model was tested in the 4 following conditions: 1) intact, 2) lateral cage alone, 3) lateral cage and plate 4) anterior cage and plate. Results were analysed using ANOVA with post-hoc Tukey’s HSD test. Range of motion (ROM) exhibited logarithmic growth with cycle number (increases of 16%, 37.5% and 24.3% in AR, FE and LB respectively). No signification difference (p > 0.1) was detected between 4 cycles, 10000 cycles and 12 hour rest stages. All measured parameters were comparable to that of reported cadaveric values. The ROM for a lateral cage and plate construct was not significantly different to the anterior lumbar interbody construct for FE (p = 1.00), LB (p = 0.995) and AR (p = 0.837). Based on anatomical and biomechanical similarities, the synthetic spine tested here provides a reasonable model to represent the human lumbar spine. Repeated testing did not dramatically alter biomechanics which may allow non-destructive testing between many different procedures and devices without the worry of carry over effects. Small intra-specimen variability and lack of biohazard makes this an attractive alternative for in vitro spine biomechanical testing. It also proved an acceptable surrogate for biomechanical testing, confirming that a lateral lumbar interbody cage and plate construct reduces ROM to a similar degree as anterior lumbar interbody cage and plate constructs.Keywords
This publication has 33 references indexed in Scilit:
- Biomechanical Analysis and Review of Lateral Lumbar Fusion ConstructsSpine, 2010
- Effect of multiple freeze–thaw cycles on intervertebral dynamic motion characteristics in the porcine lumbar spineJournal of Biomechanics, 2008
- Limitations of the Cervical Porcine Spine in Evaluating Spinal Implants in Comparison With Human Cervical Spinal SegmentsSpine, 2005
- Biomechanical comparison: stability of lateral-approach anterior lumbar interbody fusion and lateral fixation compared with anterior-approach anterior lumbar interbody fusion and posterior fixation in the lower lumbar spineJournal of Neurosurgery: Spine, 2005
- Lumbar lateral interbody cage with plate augmentation: in vitro biomechanical analysisEuropean Spine Journal, 2001
- An improved method for measuring tibiofemoral contact areas in total knee arthroplasty: a comparison of K-scan sensor and Fuji filmJournal of Biomechanics, 1999
- Are Sheep Spines a Valid Biomechanical Model for Human Spines?Spine, 1997
- A Mechanical Model of Human Spinal Motion Segments - Ein mechanisches Modell für humane WirbelsäulenbewegungssegmenteBiomedical Engineering / Biomedizinische Technik, 1997
- Stability Increase of the Lumbar Spine With Different Muscle GroupsSpine, 1995
- Human Lumbar VertebraeSpine, 1992