Review and investigation of automatic brightness/dose rate control logic of fluoroscopic imaging systems in cardiovascular interventional angiography
- 20 January 2022
- journal article
- review article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 15 (1) , 6-24
Abstract
In this article, we review automatic brightness control (ABC) for fluoroscopy imaging systems. Starting from the simple manual control, the discussion is extended to the kV-primary ABC system, and then to the most recent contrast-to-noise ratio optimized (CNR Optimized) automatic dose rate control system (ADRC). The nature of this review article is trifold. First, it describes the ABC/ADRC and associated circuits governing the operation of the fluoroscopy imaging chain. Second, we show the characteristics of a control logic from a radiation physics point of view. Third, we introduce the most recent activities in the evaluation of CNR-optimized fluoroscopy systems and the phantom design that would be compatible with the design concept of the ADRC. Because of these three subject items in the discussion process, this article is also educational in nature written for medical physicists and radiological technologists who might be less familiar with the design concept of fluoroscopy operation, specifically on the ABC and ADRC. We insert a few related matters associated with fluoroscopy automatic control circuits where they seem applicable and appropriate to enhance the understanding of fluoroscopy operation logic.Keywords
This publication has 10 references indexed in Scilit:
- Comparison of a contrast‐to‐noise ratio‐driven exposure control and a regular detector dose‐driven exposure control in abdominal imaging in a clinical angiography systemMedical Physics, 2021
- Impact of a contrast-to-noise ratio driven and material specific exposure control on image quality and radiation exposure in angiographyPhysics in Medicine & Biology, 2021
- Signal and contrast to noise ratio evaluation of fluoroscopic loops for interventional fluoroscope quality controlJournal of Applied Clinical Medical Physics, 2019
- Implementation of a spatio-temporal figure of merit for new automatic dose rate control regimes in dynamic x-ray imagingPhysics in Medicine & Biology, 2019
- Evaluation of automatic dose rate control for flat panel imaging using a spatial frequency domain figure of meritPhysics in Medicine & Biology, 2017
- Functionality and Operation of Fluoroscopic Automatic Brightness Control/Automatic Dose Rate Control Logic in Modern Cardiovascular and Interventional Angiography Systems2012
- AAPM/RSNA Physics Tutorial for Residents: Physics of Flat-Panel Fluoroscopy SystemsRadioGraphics, 2011
- The operation logic of automatic dose control of fluoroscopy system in conjunction with spectral shaping filtersMedical Physics, 2007
- The AAPM/RSNA Physics Tutorial for Residents General Overview of Fluoroscopic ImagingRadioGraphics, 2000
- Performance Evaluation and Quality Assurance in Digital Subtraction Angiography1985