Design and Evaluation of a New General-Purpose Device for Calibrating Instrumented Spatial Linkages
Authors:
- 7 January 2009
- journal article
- research article
- Published by Wiley in Journal of Biomechanical Engineering
- Vol. 131 (3) , 034505
Abstract
Because instrumented spatial linkages (ISLs) have been commonly used in measuring joint rotations and must be calibrated before using the device in confidence, a calibration device design and associated method for quantifying calibration device error would be useful. The objectives of the work reported by this paper were to (1) design an ISL calibration device and demonstrate the design for a specific application, (2) describe a new method for calibrating the device that minimizes measurement error, and (3) quantify measurement error of the device using the new method. Relative translations and orientations of the device were calculated via a series of transformation matrices containing inherent fixed and variable parameters. These translations and orientations were verified with a coordinate measurement machine, which served as a gold standard. Inherent fixed parameters of the device were optimized to minimize measurement error. After parameter optimization, accuracy was determined. The root mean squared error (RMSE) was 0.175 deg for orientation and 0.587 mm for position. All RMSE values were less than 0.8% of their respective full-scale ranges. These errors are comparable to published measurement errors of ISLs for positions and lower by at least a factor of 2 for orientations. These errors are in spite of the many steps taken in design and manufacturing to achieve high accuracy. Because it is challenging to achieve the accuracy required for a custom calibration device to serve as a viable gold standard, it is important to verify that a calibration device provides sufficient precision to calibrate an ISL.Keywords
This publication has 25 references indexed in Scilit:
- Design and Demonstration of a New Instrumented Spatial Linkage for Use in a Dynamic Environment: Application to Measurement of Ankle Rotations During SnowboardingJournal of Biomechanical Engineering, 2006
- On improving the accuracy of instrumented spatial linkage systemJournal of Biomechanics, 1996
- 6R Instrumented Spatial Linkages for Anatomical Joint Motion Measurement—Part 2: CalibrationJournal of Biomechanical Engineering, 1992
- 6R Instrumented Spatial Linkages for Anatomical Joint Motion Measurement—Part 1: DesignJournal of Biomechanical Engineering, 1992
- The Three-Dimensional Kinematics and Flexibility Characteristics of the Human Ankle and Subtalar Joint—Part II: Flexibility CharacteristicsJournal of Biomechanical Engineering, 1988
- The Three-Dimensional Kinematics and Flexibility Characteristics of the Human Ankle and Subtalar Joints—Part I: KinematicsJournal of Biomechanical Engineering, 1988
- Description and Error Evaluation of an In Vitro Knee Joint Testing SystemJournal of Biomechanical Engineering, 1988
- A Technique for Kinematic Modeling of Anatomical JointsJournal of Biomechanical Engineering, 1980
- Justification of triaxial goniometer for the measurement of joint rotationJournal of Biomechanics, 1980
- An Iterative Method for the Displacement Analysis of Spatial MechanismsJournal of Applied Mechanics, 1964