Accurate Height and Volume Measurements on Soft Samples with the Atomic Force Microscope
- 8 October 2004
- journal article
- research article
- Published by American Chemical Society (ACS) in Langmuir
- Vol. 20 (23) , 10038-10045
Abstract
The suitability of three common atomic force microscope (AFM) imaging modes for quantitative height and volume measurements on soft samples was investigated. The height and volume of rehydrated human metaphase chromosomes in liquid were measured using the contact mode, the tapping mode, and the force mapping mode. In both the contact and tapping modes, the measured height and volume strongly depended on the imaging setpoint that sets the imaging force. Measurement deviations up to 50% were observed. The force mapping mode, on the other hand, yielded reproducible height and volume measurements independent of the imaging force. It is therefore suggested that the force mapping mode should be used whenever the height or volume of soft samples need to be accurately determined.Keywords
This publication has 44 references indexed in Scilit:
- Dynamic interactions of p53 with DNA in solution by time-lapse atomic force microscopyJournal of Molecular Biology, 2001
- GTG banding pattern on human metaphase chromosomes revealed by high resolution atomic‐force microscopyJournal of Microscopy, 2001
- Drug-Induced Changes of Cytoskeletal Structure and Mechanics in Fibroblasts: An Atomic Force Microscopy StudyBiophysical Journal, 2000
- Selective Cleaning of the Cell Debris in Human Chromosome Preparations Studied by Scanning Force MicroscopyJournal of Structural Biology, 1999
- Electrostatically Balanced Subnanometer Imaging of Biological Specimens by Atomic Force MicroscopeBiophysical Journal, 1999
- Chromatin Structure in Bands and Interbands of Polytene Chromosomes Imaged by Atomic Force MicroscopyJournal of Structural Biology, 1998
- Short cantilevers for atomic force microscopyReview of Scientific Instruments, 1996
- Chromosome classification by atomic force microscopy volume measurementJournal of Vacuum Science & Technology B: Microelectronics and Nanometer Structures Processing, Measurement, and Phenomena, 1996
- Chromosomes are highly elastic and can be stretchedCytogenetic and Genome Research, 1994
- The relation between load and penetration in the axisymmetric boussinesq problem for a punch of arbitrary profileInternational Journal of Engineering Science, 1965