Downloadable Software Algorithm Reduces Inappropriate Shocks Caused by Implantable Cardioverter-Defibrillator Lead Fractures
- 12 October 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 122 (15) , 1449-1455
Abstract
Background—: Downloadable software upgrades are common in consumer electronics but not in implantable medical devices. Fractures of implantable cardioverter-defibrillator (ICD) leads present commonly as inappropriate shocks. A lead-integrity alert (LIA) designed to reduce inappropriate shocks is the first software download approved to enhance nominally functioning, previously implanted ICDs. Methods and Results—: We performed a prospective study to determine whether an LIA could reduce inappropriate shocks. Patients were included if they had ICD lead fractures confirmed by analysis of explanted leads. The LIA group included the first 213 patients who met the inclusion criteria after the LIA was approved who had the LIA downloaded. The LIA is triggered either by high impedance or rapid oversensing. It responds by delaying detection of ventricular fibrillation and initiating a patient alert every 4 hours. The control group included the first 213 patients who did not have the LIA downloaded. They were monitored by conventional daily impedance measurements that respond with a daily alert. The LIA group had a 46% relative reduction (95% confidence interval 34% to 55%) in the percentage of patients with ≥1 inappropriate shock (LIA 38% versus control 70%, P P P <0.001). Conclusions—: A software download that upgrades previously implanted ICDs without surgical revision reduces inappropriate shocks caused by lead fractures.Keywords
This publication has 25 references indexed in Scilit:
- Limitation of Programmed Alerts to Predict ICD Lead FailuresPacing and Clinical Electrophysiology, 2009
- Analysis of Pacing/Defibrillator Lead Failure Using Device Diagnostics and Pacing ManeuversPacing and Clinical Electrophysiology, 2009
- Multiple ICD Discharges Associated with Lead Fracture Without Triggering of High Impedance AlertPacing and Clinical Electrophysiology, 2009
- Downloadable Algorithm to Reduce Inappropriate Shocks Caused by Fractures of Implantable Cardioverter-Defibrillator LeadsCirculation, 2008
- Failure of impedance monitoring to prevent adverse clinical events caused by fracture of a recalled high-voltage implantable cardioverter-defibrillator leadHeart Rhythm, 2008
- Outcome of the Fidelis implantable cardioverter-defibrillator lead advisory: A report from the Canadian Heart Rhythm Society Device Advisory CommitteeHeart Rhythm, 2008
- When Life‐Saving Devices Terminate LifeJournal of Cardiovascular Electrophysiology, 2008
- Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator leadHeart Rhythm, 2007
- Annual Rate of Transvenous Defibrillation Lead Defects in Implantable Cardioverter-Defibrillators Over a Period of >10 YearsCirculation, 2007
- An algorithm to predict implantable cardioverter-defibrillator lead failureJACC, 2004