Stiffness Characteristics of Composite Hybrid External Fixators
Authors:
- Roland GlöcknerRGRoland Glöckner
- Hermann BailHBHermann Bail
- Stefan KolbeckSKStefan Kolbeck
- Michael RaschkeMRMichael Raschke
- 1 December 2002
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 405 (405) , 267-276
Abstract
New composite hybrid fixators for fracture stabilization using single or multiple rings with monolateral bars and half-pin fixation may provide clinical advantages such as reduced bulk and easier skin access. However, the mechanical properties of these fixators are difficult to estimate from numerous design parameters. Addressing this problem the following research questions were asked: (1) Do composite hybrid fixators have similar stiffness properties to clinically proven triple-ring fixators; (2) How does the fixation mechanism affect the stiffness properties of external fixators, and, (3) Are there interferences between ring number and fixation method? An experimental study was done on simulated metaphyseal tibial fractures and stiffness of 12 fixators was measured for different loading conditions. The results showed that triple-ring fixators provide approximately 20% stiffer properties than double-ring and single-ring fixators. No influence of ring number on the AP bending properties was found and different fixation methods were associated with large differences in fixator stiffness, whereas significant interferences were found between ring number and fixation method. Although the mechanical properties investigated in the current study cannot predict directly the clinical performance of these fixators, the stiffness data provide useful information for making decisions regarding the treatment of fractures using external fixation.Keywords
This publication has 13 references indexed in Scilit:
- The effects of different wire and screw combinations on the stiffness of a hybrid external fixatorProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2000
- Biomechanical Comparison of Hybrid External FixatorsJournal of Orthopaedic Trauma, 1998
- Supportive Composite- Hybridfixation perkutan verschraubter TibiakopffrakturenDer Unfallchirurg, 1996
- Comparative Biomechanical Evaluation of Different External Fixation Sidebars: Stainless-Steel Tubes Versus Carbon Fiber RodsJournal of Orthopaedic Trauma, 1996
- Influences of Some Design Parameters on the Biomechanics of the Unreamed Tibial Intramedullary NailJournal of Orthopaedic Trauma, 1993
- Basic biomechanical measurements of bone: A tutorialBone, 1993
- Stiffness Characteristics of the Circular Ilizarov Device as Opposed to Conventional External FixatorsJournal of Biomechanical Engineering, 1990
- The Tension-Stress Effect on the Genesis and Growth of TissuesClinical Orthopaedics and Related Research, 1989
- The comparative stiffness of external fixation framesInjury, 1981
- How to increase the stability of external fixation unitsArchives of orthopaedic and trauma surgery, 1980