Improved assessment of cartilage repair tissue using fluid-suppressed 23Na inversion recovery MRI at 7 Tesla: preliminary results
- 18 February 2012
- journal article
- musculoskeletal
- Published by Springer Nature in European Radiology
- Vol. 22 (6) , 1341-1349
Abstract
Objectives To evaluate cartilage repair and native tissue using a three-dimensional (3D), radial, ultra-short echo time (UTE) 23Na MR sequence without and with an inversion recovery (IR) preparation pulse for fluid suppression at 7 Tesla (T). Methods This study had institutional review board approval. We recruited 11 consecutive patients (41.5 ± 11.8 years) from an orthopaedic surgery practice who had undergone a knee cartilage restoration procedure. The subjects were examined postoperatively (median = 26 weeks) with 7-T MRI using: proton-T2 (TR/TE = 3,000 ms/60 ms); sodium UTE (TR/TE = 100 ms/0.4 ms); fluid-suppressed, sodium UTE adiabatic IR. Cartilage sodium concentrations in repair tissue ([Na+]R), adjacent native cartilage ([Na+]N), and native cartilage within the opposite, non-surgical compartment ([Na+]N2) were calculated using external NaCl phantoms. Results For conventional sodium imaging, mean [Na+]R, [Na+]N, [Na+]N2 were 177.8 ± 54.1 mM, 170.1 ± 40.7 mM, 172.2 ± 30 mM respectively. Differences in [Na+]R versus [Na+]N (P = 0.59) and [Na+]N versus [Na+]N2 (P = 0.89) were not significant. For sodium IR imaging, mean [Na+]R, [Na+]N, [Na+]N2 were 108.9 ± 29.8 mM, 204.6 ± 34.7 mM, 249.9 ± 44.6 mM respectively. Decreases in [Na+]R versus [Na+]N (P = 0.0.0000035) and [Na+]N versus [Na+]N2 (P = 0.015) were significant. Conclusions Sodium IR imaging at 7 T can suppress the signal from free sodium within synovial fluid. This may allow improved assessment of [Na+] within cartilage repair and native tissue. Key Points • NaIR magnetic resonance imaging can suppress signal from sodium within synovial fluid. • NaIR MRI thus allows assessment of sodium concentration within cartilage tissue alone. • This may facilitate more accurate assessment of repair tissue composition and quality.Keywords
This publication has 32 references indexed in Scilit:
- Sodium relaxation times in the knee joint in vivo at 7TNMR in Biomedicine, 2011
- Sodium inversion recovery MRI of the knee joint in vivo at 7TJournal of Magnetic Resonance, 2010
- In vivo sodium imaging of human patellar cartilage with a 3D cones sequence at 3 T and 7 TJournal of Magnetic Resonance Imaging, 2010
- Rapid isotropic 3D‐sodium MRI of the knee joint in vivo at 7TJournal of Magnetic Resonance Imaging, 2009
- Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer (gagCEST)Proceedings of the National Academy of Sciences, 2008
- In vivo T1ρ and T2 mapping of articular cartilage in osteoarthritis of the knee using 3 T MRIOsteoarthritis and Cartilage, 2007
- Quantitative T2 Mapping of Matrix-Associated Autologous Chondrocyte Transplantation at 3 TeslaInvestigative Radiology, 2007
- Sodium and T1ρ MRI for molecular and diagnostic imaging of articular cartilageNMR in Biomedicine, 2006
- Global and regional reproducibility of T2 relaxation time measurements in human patellar cartilageMagnetic Resonance in Medicine, 2006
- Magnetic Resonance Imaging of Articular CartilageThe American Journal of Sports Medicine, 2006