Rounded leaf end modeling in Pinnacle VMAT treatment planning for fixed jaw linacs
Open Access
- 8 November 2016
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 17 (6) , 149-162
Abstract
During volume-modulated arc therapies (VMAT), dosimetric errors are introduced by multiple open dynamic leaf gaps that are present in fixed diaphragm linear accelerators. The purpose of this work was to develop a methodology for adjusting the rounded leaf end modeling parameters to improve out-of-field dose agreement in SmartArc VMAT treatment plans delivered by fixed jaw linacs where leaf gap dose is not negligible. Leaf gap doses were measured for an Elekta beam modulator linac with 0.4 cm micro-multileaf collimators (MLC) using an A16 micro-ionization chamber, a MatriXX ion chamber detector array, and Kodak EDR2 film dosimetry in a solid water phantom. The MLC offset and rounded end tip radius were adjusted in the Pinnacle treatment planning system (TPS) to iteratively arrive at the optimal configuration for 6 MV and 10 MV photon energies. Improvements in gamma index with a acceptance criteria and an inclusion threshold of 5% of maximum dose were measured, analyzed, and validated using an ArcCHECK diode detector array for field sizes ranging from 1.6 to 14 cm square field arcs and Task Group (TG) 119 VMAT test cases. The best results were achieved for a rounded leaf tip radius of 13 cm with a 0.1 cm MLC offset. With the optimized MLC model, measured gamma indices ranged between 99.9% and 91.7% for square field arcs with sizes between 3.6 cm and 1.6 cm, with a maximum improvement of 42.7% for the 1.6 cm square field size. Gamma indices improved up to 2.8% in TG-119 VMAT treatment plans. Imaging and Radiation Oncology Core (IROC) credentialing of a VMAT plan with the head and neck phantom passed with a gamma index of 100%. Fine-tune adjustments to MLC rounded leaf ends may improve patient-specific QA pass rates and provide more accurate predictions of dose deposition to avoidance structures. PACS number(s): 87.55.D-, 87.55.kd, 87.55.kh
Keywords
This publication has 42 references indexed in Scilit:
- Direct impact analysis of multi-leaf collimator leaf position errors on dose distributions in volumetric modulated arc therapy: a pass rate calculation between measured planar doses with and without the position errorsPhysics in Medicine & Biology, 2011
- Intensity-modulated arc therapy: principles, technologies and clinical implementationPhysics in Medicine & Biology, 2011
- Optimal sensitometric curves of Kodak EDR2 film for dynamic intensity modulated radiation therapy verificationBiomedical Imaging and Intervention Journal, 2008
- Multileaf collimator end leaf leakage: implications for wide-field IMRTPhysics in Medicine & Biology, 2007
- Intensity modulated radiation therapy with irregular multileaf collimated field: A dosimetric study on the penumbra region with different leaf stepping patternsMedical Physics, 2006
- Verification of a rounded leaf-end MLC model used in a radiotherapy treatment planning systemPhysics in Medicine & Biology, 2006
- Elements of commissioning step-and-shoot IMRT: Delivery equipment and planning system issues posed by small segment dimensions and small monitor unitsMedical Dosimetry, 2005
- Dosimetric characteristics of the Elekta Beam Modulator™Physics in Medicine & Biology, 2005
- Lateral loss and dose discrepancies of multileaf collimator segments in intensity modulated radiation therapyMedical Physics, 2003
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998