Direct impact analysis of multi-leaf collimator leaf position errors on dose distributions in volumetric modulated arc therapy: a pass rate calculation between measured planar doses with and without the position errors
Authors:
- D TatsumiDTD Tatsumi
- M N HosonoMHM N Hosono
- R NakadaRNR Nakada
- K IshiiKIK Ishii
- S TsutsumiSTS Tsutsumi
- M InoueMIM Inoue
- T IchidaTIT Ichida
- Y MikiYMY Miki
- 30 September 2011
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (20) , N237-N246
Abstract
We propose a new method for analyzing the direct impact of multi-leaf collimator (MLC) leaf position errors on dose distributions in volumetric modulated arc therapy (VMAT). The technique makes use of the following processes. Systematic leaf position errors are generated by directly changing a leaf offset in a linac controller; dose distributions are measured by a two-dimensional diode array; pass rates of the dose difference between measured planar doses with and without the position errors are calculated as a function of the leaf position error. Three different treatment planning systems (TPSs) were employed to create VMAT plans for five prostate cancer cases and the pass rates were compared between the TPSs under various leaf position errors. The impact of the leaf position errors on dose distributions depended upon the final optimization result from each TPS, which was explained by the correlation between the dose error and the average leaf gap width. The presented method determines leaf position tolerances for VMAT delivery for each TPS, which may facilitate establishing a VMAT quality assurance program in a radiotherapy facility.Keywords
This publication has 12 references indexed in Scilit:
- Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapyRadiotherapy and Oncology, 2010
- Tolerances on MLC leaf position accuracy for IMRT delivery with a dynamic MLCMedical Physics, 2009
- Development and evaluation of an efficient approach to volumetric arc therapy planningMedical Physics, 2009
- Quality assurance of volumetric modulated arc therapy using Elekta SynergyActa Oncologica, 2009
- Volumetric modulated arc therapy: IMRT in a single gantry arcMedical Physics, 2007
- Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancerPhysics in Medicine & Biology, 2007
- An accurate calibration method of the multileaf collimator valid for conformal and intensity modulated radiation treatmentsPhysics in Medicine & Biology, 2004
- A 2-D diode array and analysis software for verification of intensity modulated radiation therapy deliveryMedical Physics, 2003
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998
- The design and performance characteristics of a multileaf collimatorPhysics in Medicine & Biology, 1994