Radiation risk assessment in neonatal radiographic examinations of the chest and abdomen: a clinical and Monte Carlo dosimetry study
Authors:
- E YakoumakisEYE Yakoumakis
- D PapadopoulouDPD Papadopoulou
- G GialousisGGG Gialousis
- V TheodoropoulosVTV Theodoropoulos
- P SandilosPSP Sandilos
- E GeorgiouEGE Georgiou
- 19 September 2006
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (19) , 5023-5033
Abstract
Seeking to assess the radiation risk associated with radiological examinations in neonatal intensive care units, thermo-luminescence dosimetry was used for the measurement of entrance surface dose (ESD) in 44 AP chest and 28 AP combined chest-abdominal exposures of a sample of 60 neonates. The mean values of ESD were found to be equal to 44 +/- 16 microGy and 43 +/- 19 microGy, respectively. The MCNP-4C2 code with a mathematical phantom simulating a neonate and appropriate x-ray energy spectra were employed for the simulation of the AP chest and AP combined chest-abdominal exposures. Equivalent organ dose per unit ESD and energy imparted per unit ESD calculations are presented in tabular form. Combined with ESD measurements, these calculations yield an effective dose of 10.2 +/- 3.7 microSv, regardless of sex, and an imparted energy of 18.5 +/- 6.7 microJ for the chest radiograph. The corresponding results for the combined chest-abdominal examination are 14.7 +/- 7.6 microSv (males)/17.2 +/- 7.6 microSv (females) and 29.7 +/- 13.2 microJ. The calculated total risk per radiograph was low, ranging between 1.7 and 2.9 per million neonates, per film, and being slightly higher for females. Results of this study are in good agreement with previous studies, especially in view of the diversity met in the calculation methods.Keywords
This publication has 30 references indexed in Scilit:
- Differences in effective dose and energy imparted estimation from PA–AP, RLAT–LLAT projections in pediatric full spine x-ray examination using the Monte Carlo techniquePhysics in Medicine & Biology, 2005
- Radiation dose to premature infants in neonatal intensive care units in kuwaitRadiation Protection Dosimetry, 2004
- A guide to GUMEuropean Journal of Physics, 2002
- Effective dose and energy imparted in diagnostic radiologyMedical Physics, 1997
- Radiology in the neonatal intensive care unit: dose reduction and image qualityBritish Journal of Radiology, 1996
- An investigation into techniques for reducing doses from neo-natal radiographic examinationsBritish Journal of Radiology, 1995
- Radiation dose to neonates on a Special Care Baby UnitBritish Journal of Radiology, 1989
- The risk of diagnostic radiation of the newbornBritish Journal of Radiology, 1986
- Radiation doses to neonates requiring intensive careBritish Journal of Radiology, 1983
- Active bone marrow distribution as a function of age in humansPhysics in Medicine & Biology, 1981