Abstract
Entrance surface dose (ESD) and effective dose (E) to premature infants were estimated at three neonatal intensive care units in Kuwait for three standard X-ray examinations—abdominal, chest and skull X rays using a simple water phantom. The ESD was found to vary between 58 and 102 μGy for abdominal X rays, between 51 and 102 μGy for chest X rays and between 58 and 145 μGy for skull examinations. These doses are comparable to the entrance skin doses published elsewhere. The E-values were estimated using normalised organ dose dataset from the National Radiological Protection Board. The E-values for abdominal, chest and skull examinations were in the ranges of 30–46, 20–36 and 8–18 μSv per examination, respectively. The risk of developing childhood cancers from each of the three examinations was estimated to be in the range (9–117) × 10−6 for infants undergoing 25 of these X-ray examinations during their stay in the NIC unit.