Abstract
Objective:To evaluate the impact of a low-dose (LD) and an ultra-LD (ULD) contrast protocol for coronary CT angiography on qualitative and quantitative image parameters in a clinical setting.Methods:We scanned 120 consecutive patients with a 256-slice CT scanner applying a LD (60 patients, 35–55 ml) or ULD (60 patients, 25–45 ml) contrast protocol adapted to the body mass index. Visually assessed image quality and attenuation measured in each coronary segment were retrospectively compared in 20 consecutive patients scanned with a normal-dose (ND, 40–105 ml) contrast protocol.Results:Visually assessed image quality did not differ significantly among protocols. By contrast, attenuation obtained from the ULD protocol (median contrast volume 35 ml) differed significantly from the LD (median 45 ml) and ND (median 70 ml) protocols in the coronary segments (316 ± 52 vs 363 ± 60 and 359 ± 52 HU, p < 0.001). Attenuation did not differ significantly between the LD and ND protocol. The proportion of patients with in...