Comparison of 18F SPECT with PET in myocardial imaging: A realistic thorax-cardiac phantom study
Open Access
- 31 October 2006
- journal article
- Published by Springer Nature in BMC Nuclear Medicine
- Vol. 6 (1) , 5
Abstract
Positron emission tomography (PET) imaging with fluorine-18 (18F) Fluorodeoxyglucose (FDG) and flow tracer such as Rubidium-82 (82Rb) is an established method for evaluating an ischemic but viable myocardium. However, the high cost of PET imaging restricts its wider clinical use. Therefore, less expensive 18F FDG single photon emission computed tomography (SPECT) imaging has been considered as an alternative to 18F FDG PET imaging. The purpose of the work is to compare SPECT with PET in myocardial perfusion/viability imaging. A nonuniform RH-2 thorax-heart phantom was used in the SPECT and PET acquisitions. Three inserts, 3 cm, 2 cm and 1 cm in diameter, were placed in the left ventricular (LV) wall to simulate infarcts. The phantom acquisition was performed sequentially with 7.4 MBq of 18F and 22.2 MBq of Technetium-99m (99mTc) in the SPECT study and with 7.4 MBq of 18F and 370 MBq of 82Rb in the PET study. SPECT and PET data were processed using standard reconstruction software provided by vendors. Circumferential profiles of the short-axis slices, the contrast and viability of the inserts were used to evaluate the SPECT and PET images. The contrast for 3 cm, 2 cm and 1 cm inserts were for 18F PET data, 1.0 +/- 0.01, 0.67 +/- 0.02 and 0.25 +/- 0.01, respectively. For 82Rb PET data, the corresponding contrast values were 0.61 +/- 0.02, 0.37 +/- 0.02 and 0.19 +/- 0.01, respectively. For 18F SPECT the contrast values were, 0.31 +/- 0.03 and 0.20 +/- 0.05 for 3 cm and 2 cm inserts, respectively. For 99mTc SPECT the contrast values were, 0.63 +/- 0.04 and 0.24 +/- 0.05 for 3 cm and 2 cm inserts respectively. In SPECT, the 1 cm insert was not detectable. In the SPECT study, all three inserts were falsely diagnosed as "viable", while in the PET study, only the 1 cm insert was diagnosed falsely "viable". For smaller defects the 99mTc/18F SPECT imaging cannot entirely replace the more expensive 82Rb/18F PET for myocardial perfusion/viability imaging, due to poorer image spatial resolution and poorer defect contrast.Keywords
This publication has 24 references indexed in Scilit:
- Phantom studies for estimation of defect size on cardiac (18)F SPECT and PET: implications for myocardial viability assessment.2001
- Blood flow–metabolism imaging with positron emission tomography in patients with diabetes mellitus for the assessment of reversible left ventricular contractile dysfunction 11The Laboratory of Structural Biology and Molecular Medicine is operated for the U.S. Department of Energy by the University of California under Contract #DE-FC03-87ER60615.JACC, 1999
- Functional assessment with electrocardiographic gated single-photon emission computed tomography improves the ability of technetium-99m sestamibi myocardial perfusion imaging to predict myocardial viability in patients undergoing revascularizationThe American Journal of Cardiology, 1999
- Fluorine-18-fluorodeoxyglucose cardiac imaging using a modified scintillation camera.1998
- Delay in revascularization is associated with increased mortality rate in patients with severe left ventricular dysfunction and viable myocardium on fluorine 18-fluorodeoxyglucose positron emission tomography imaging.1998
- Predictive Value of Planar 1 8F-Fluorodeoxyglucose Imaging for Cardiac Events in Patients After Acute Myocardial InfarctionThe American Journal of Cardiology, 1998
- Preoperative Positron Emission Tomographic Viability Assessment and Perioperative and Postoperative Risk in Patients With Advanced Ischemic Heart DiseaseJACC, 1997
- Three-window transformation cross-talk correction for simultaneous dual-isotope imaging.1997
- High-energy (511-keV) imaging with the scintillation camera.RadioGraphics, 1996
- MYOCARDIAL PERFUSION WITH RUBIDIUM-82 .1. MEASUREMENT OF EXTRACTION FRACTION AND FLOW WITH EXTERNAL DETECTORS1983