Objectives: This study examined the emission characteristics of ethylene oxide (EO)-sterilized syringes under various environmental conditions, aiming to develop control strategies to minimize worker exposure. Methods: Experiments were performed in a facility in which temperature, relative humidity (RH), and air change rate (ACR) were controlled. Results: Analytical results indicate that the main effects of the four test variables on kinetic parameters were statistically significant (p < 0.05), except for the effect of the product on the decay rate constant, the effect of ACR on maximum EO concentration, and effect of RH on the area under the curve-days 1 and 2. The interactive effects among test variables were also evident, indicating complex emission behaviors. The mean EO emission factors during the days 1 and 2 and at the 48th hour for the 1- and 30-ml products were 2302, 1301, and 1031 mg/m3/h, and 871, 490, and 381 mg/m3/h, respectively. The times required for air EO concentrations from tested products to return to ∼0 and 1 ppm (permissible limit) were 417 and 218 h, respectively. Conclusions: Plastic content, temperature, RH, and ACR affected EO emissions. ACR is an achievable means of control; however, the aeration area/system should be isolated to ensure adequate ventilation is achieved. © 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2009Keywords
This publication has 20 references indexed in Scilit:
- Ethylene oxide sterilization in the medical‐supply manufacturing industry: Assessment and control of worker exposureJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2007
- Carcinogenicity and Genotoxicity of Ethylene Oxide: New Aspects and Recent AdvancesCritical Reviews in Toxicology, 2000
- Considerations on Revisions of Emissions Testing Protocols1996
- Determination of Worker Exposure to Instantaneously Emitted Ethylene Oxide in Sterilization FacilitiesApplied Industrial Hygiene, 1989
- Computerized model for accurate determination of ethylene oxide diffusion in sterilized medical suppliesBiomaterials, 1989
- Temperature and humidity dependence of formaldehyde release from selected building materialsBulletin of Environmental Contamination and Toxicology, 1989
- Aeration of ethylene oxide-sterilized polymersBiomaterials, 1986
- Ethylene oxide sterilization: effect of polymer structure and sterilization conditions on residue levelsBiomaterials, 1980
- Ethylene Oxide Degassing of Rubber and Plastic MaterialsAnesthesiology, 1976
- Aeration of medical plasticsJournal of Biomedical Materials Research, 1974