A Review of the First Long-term Implantable Continuous Glucose Monitoring System Available in the United States
Author:
- 13 December 2019
- journal article
- review article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 15 (1) , 160-166
Abstract
Background: Although real-time continuous glucose monitoring (rtCGM) has been shown to improve glycemic control in patients with type 1 diabetes mellitus and type 2 diabetes mellitus treated with insulin, rates of adoption have been low. A novel approach, with the use of a long-term implantable continuous glucose monitoring (LTI CGM) has the potential to overcome barriers to rtCGM. The purpose of this review is to provide a background on the first LTI CGM technology to be approved, along with a review of contraindications, interference, safety, accuracy, and efficacy. Considerations for patient selection are discussed based on the available evidence. Methods: PubMed, EMBASE, and Cochrane Library were searched for keywords and subject headings to identify studies assessing LTI CGM. Results: Seven studies were identified which assessed LTI CGM. Mean absolute relative difference is similar to available CGM devices. Rates of adverse events were low. Change in hemoglobin A1c with LTI CGM may be comparable to rtCGM. Conclusions: Based on the available evidence, LTI CGM appears to be safe and accurate. Additional clinical trial investigation is warranted to evaluate the glycemic efficacy of LTI CGM.Keywords
This publication has 22 references indexed in Scilit:
- A Prospective Multicenter Evaluation of the Accuracy and Safety of an Implanted Continuous Glucose Sensor: The PRECISION StudyDiabetes Technology & Therapeutics, 2019
- Reference Guide for Integrating Continuous Glucose Monitoring Into Clinical PracticeThe Diabetes Educator, 2018
- Interference Assessment of Various Endogenous and Exogenous Substances on the Performance of the Eversense Long-Term Implantable Continuous Glucose Monitoring SystemDiabetes Technology & Therapeutics, 2018
- A Prospective Multicenter Evaluation of the Accuracy of a Novel Implanted Continuous Glucose Sensor: PRECISE IIDiabetes Technology & Therapeutics, 2018
- Continuous Glucose Monitoring Versus Usual Care in Patients With Type 2 Diabetes Receiving Multiple Daily Insulin InjectionsAnnals of Internal Medicine, 2017
- Effect of Continuous Glucose Monitoring on Glycemic Control in Adults With Type 1 Diabetes Using Insulin InjectionsJAMA, 2017
- Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled TrialDiabetes Therapy, 2016
- Accuracy and Longevity of an Implantable Continuous Glucose Sensor in the PRECISE Study: A 180-Day, Prospective, Multicenter, Pivotal TrialDiabetes Care, 2016
- Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trialThe Lancet, 2016
- Real-Time Continuous Glucose Monitoring Among Participants in the T1D Exchange Clinic RegistryDiabetes Care, 2014