Solid-state electrolyte sensors for rebreather applications: a preliminary investigation

Authors:

  • 1 June 2011
    • journal article
    • research article
Abstract
(Sieber A, Baumann R, Fasoulas S, Krozer A. Solid-state electrolyte sensors for rebreather applications: a preliminary investigation. Diving and Hyperbaric Medicine. 2011;41 (2):90-6.) Introduction: Recently developed prototypes of zirconium dioxide and NASICON-based micro solid-state electrolyte oxygen (O(2)) and carbon dioxide (CO(2)) sensors were tested for their potential suitability in rebreathers. The O(2) sensor has a quasi-indefinite lifetime, whilst that of the CO(2) sensor is approximately 700 h. This is a preliminary report of a new technological application. Methods: The O(2) sensor was tested in a small pressure chamber to a partial pressure of oxygen (PO(2)) of 405 kPa (4 bar). The CO, sensor was tested up to 10 kPa CO(2). The response times to a step change of pressure were measured, and cross-sensitivity for helium tested using trimix. A rebreather mouthpiece was modified so that breath-by-breath gas recordings could be observed. Power consumption to heat the sensors was measured. Results: The O(2) sensor demonstrated non-linearity, particularly above 101.3 kPa (1 bar) PO(2), whereas the output of the CO(2) sensor showed an inverse logarithmic relationship. Cross-sensitivity to helium was observed. The mean t90 response times were 90 (SD 10) ms for the O(2) sensor, and 100 (SD 10) ms for the CO(2) sensor. Breath-by-breath recordings showed slight damping of the CO(2) trace due to electronic filtering. Power consumption was 1.5-2 W per sensor. Conclusions: The fast response times would allow accurate breath-by-breath measurement. Even though the O(2) sensor has a non-linear response, measurement is possible using multi-point calibration. Further design is necessary to allow trimix to be used as the diluent. A major disadvantage is the high power consumption needed to heat the sensors to high temperatures.