Determinations of organ doses and effective dose equivalents from computed tomographic examination
Authors:
- 1 January 1991
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 64 (757) , 20-28
Abstract
The organ or tissue doses were determined with a phantom measurement for 12 types of CT scanners widely used in Japan. Two types of thermoluminescent dosimeters were used for the dose determinations in a Rando woman phantom. The effective dose equivalents recommended by the International Commission on Radiological Protection were calculated using the measured organ or tissue doses. It was found that the CT scanners currently available give quite different organ or tissue doses. When selecting the optimum technical factors for scanning, therefore, it is important to take into consideration the balance of the image quality and the radiation exposure to patients.Keywords
This publication has 7 references indexed in Scilit:
- Average radiation doses in a standard head examination for 250 CT systems.Radiology, 1987
- The use of the effective dose equivalent,HE, as a risk parameter in computed tomographyBritish Journal of Radiology, 1986
- Patient doses received during whole body scanning using an Elscint 905 CT scannerBritish Journal of Radiology, 1985
- CT Dosimetry and Risk EstimatesRadiation Protection Dosimetry, 1985
- Radiation Dosimetry Descriptors Applied to Four C T Scanners in New ZealandAustralasian Radiology, 1984
- Radiation dosimetry survey of computed tomography systems from ten manufacturersBritish Journal of Radiology, 1982
- Reference Japanese Man-II. Distribution of Strontium in the Skeleton and in the Mass of Mineralized BoneHealth Physics, 1981