The 10-year outcomes of the ASR XL Acetabular System: a single-center experience from China
Open Access
- 24 May 2019
- journal article
- research article
- Published by Springer Nature in Journal of Orthopaedic Surgery and Research
- Vol. 14 (1) , 154
Abstract
The revision rate of articular surface replacement (ASR) implants continues to rise in China because of metal debris. However, there are few reports on the clinical results of ASR implants with prolonged follow-up time in China. This study investigated the clinical outcomes and the risk factors of revision surgery in patients with ASR implants. In total, 74 patients (74 hips) who underwent primary total hip arthroplasty (THA) with ASR implants over the past 4 to 10 years were retrospectively analyzed. Relevant clinical, radiographic, and biochemical data were examined. The average follow-up time was 88.46 (range 23–114) months, and the ASR implants of 18 hips (24.3%) were revised. Patients who received revision surgery had worse joint function with significantly lower Harris Hip Score and Western Ontario and McMaster Universities index than non-revision patients (61.11 ± 6.68 vs 85.30 ± 9.16, p < 0.001; 61.00 ± 3.83 vs 79.04 ± 14.49, p < 0.001; respectively). Higher acetabular abduction angle and serum Co and Cr concentration were significantly relevant to worse joint function as measured by HSS (p = 0.018, 0.009, 0.043, respectively). ROC curve analysis was applied to categorize the optimal cutoff values of acetabular abduction angle and serum Cr and Co concentration for revision surgery, which were settled as 47.80°, 98.44 μg/L, and 6.95 μg/L, respectively. Overall survival of the prostheses with high acetabular abduction angle (> 47.80°, HR = 70.145, 95% CI 1.558–3158.213, p = 0.029), high serum Cr concentration (98.44 μg/L, HR = 58.956, 95% CI 1.294–2685.203, p = 0.036), and high serum Co concentration (> 6.95 μg/L, HR = 179.511, 95% CI 2.360–13656.941, p = 0.019) decreased significantly than the lower groups. Evaluation of the DePuy ASR XL articulation demonstrated increased rates of revision following a longer follow-up period. High acetabular abduction angle and serum Cr and Co concentration correlated with worse clinical outcomes and high revision rate. Therefore, we advocate that patients with DePuy ASR XL implants be followed up more closely than those with other implants, especially with high acetabular abduction angle and serum Cr or Co concentration.Keywords
Funding Information
- Taizhou Municipal Science and Technology Bureau (15yw05)
This publication has 29 references indexed in Scilit:
- An ORS-ICP-MS method for monitoring trace levels of cobalt and chromium in whole blood samples from hip arthroplasty patients with metal-on-metal prosthesesClinical Biochemistry, 2012
- High incidence of pseudotumour formation after large-diameter metal-on-metal total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2012
- Insufficient Acetabular Version Increases Blood Metal Ion Levels after Metal-on-metal Hip ResurfacingClinical Orthopaedics and Related Research, 2011
- Adverse reaction to metal debris following hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2011
- Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2010
- Blood metal ion concentrations after hip resurfacing arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2009
- Six-year results of a prospective study of metal ion levels in young patients with metal-on-metal hip resurfacingsThe Journal of Bone and Joint Surgery. British volume, 2009
- Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacementThe Journal of Bone and Joint Surgery. British volume, 2008
- Pseudotumours associated with metal-on-metal hip resurfacingsThe Journal of Bone and Joint Surgery. British volume, 2008
- Metal-on-Metal Hip Arthroplasty Does Equally Well in Osteonecrosis and OsteoarthritisClinical Orthopaedics and Related Research, 2008