Abstract
We investigated a pinhole imaging system for independent in vivo monitoring and verification of high dose rate (HDR) brachytherapy treatment. The system consists of a high-resolution pinhole collimator, an x-ray fluoroscope, and a standard radiographic screen-film combination. Autofluoroscopy provides real-time images of the in vivo Ir-192 HDR source for monitoring the source location and movement, whereas autoradiography generates a permanent record of source positions on film. Dual-pinhole autoradiographs render stereo-shifted source images that can be used to reconstruct the source dwell positions in three dimensions. The dynamic range and spatial resolution of the system were studied with a polystyrene phantom using a range of source strengths and dwell times. For the range of source activity used in HDR brachytherapy, a 0.5 mm diameter pinhole produced sharp fluoroscopic images of the source within the dynamic range of the fluoroscope. With a source-to-film distance of 35 cm and a 400 speed screen-film combination, the same pinhole yielded well recognizable images of a 281.2 GBq (7.60 Ci) Ir-192 source for dwell times in the typical clinical range of 2 to 400 s. This 0.5 mm diameter pinhole could clearly resolve source positions separated by lateral displacements as small as 1 mm. Using a simple reconstruction algorithm, dwell positions in a phantom were derived from stereo-shifted dual-pinhole images and compared to the known positions. The agreement was better than 1 mm. A preliminary study of a patient undergoing HDR treatment for cervical cancer suggests that the imaging method is clinically feasible. Based on these studies we believe that the pinhole imaging method is capable of providing independent and reliable real-time monitoring and verification for HDR brachytherapy.Keywords
This publication has 6 references indexed in Scilit:
- High dose-rate brachytherapy treatment delivery: Report of the AAPM Radiation Therapy Committee Task Group No. 59Medical Physics, 1998
- A Monte Carlo study of verification imaging in high dose rate brachytherapyMedical Physics, 1998
- A system for nonradiographic source localization and real-time planning of intraoperative high dose rate brachytherapyMedical Physics, 1997
- The effect of gamma ray penetration on angle‐dependent sensitivity for pinhole collimation in nuclear medicineMedical Physics, 1997
- Quantitative verification of 192Ir PDR and HDR source structure by pin‐hole autoradiographyMedical Physics, 1995
- Autoradiography for iodine-125 seedsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1993