Metabolism of oxygen via T2 and interleaved velocity encoding: A rapid method to quantify whole‐brain cerebral metabolic rate of oxygen
- 14 June 2022
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 88 (3) , 1229-1243
Abstract
Purpose: Cerebral metabolic rate of oxygen (CMRO2) is an important biomarker of brain function. Key physiological parameters required to quantify CMRO2 include blood flow rate in the feeding arteries and venous oxygen saturation (SvO2) in the draining vein. Here, a pulse sequence, metabolism of oxygen via T2 and interleaved velocity encoding (MOTIVE), was developed to measure both parameters simultaneously and enable CMRO2 quantification in a single pass.Methods: The MOTIVE sequence interleaves a phase‐contrast module between a nonselective saturation and a background‐suppressed T2‐prepared EPI readout (BGS‐EPI) to measure T2 of blood water protons and cerebral blood flow in 20 s or less. The MOTIVE and standalone BGS‐EPI sequences were compared against TRUST (“T2 relaxation under spin tagging”) in the brain in healthy subjects (N = 24). Variants of MOTIVE to enhance resolution or shorten scan time were explored. Intrasession and intersession reproducibility studies were performed.Results: MOTIVE experiments yielded an average SvO2 of 61 ± 6% in the superior sagittal sinus of the brain and an average cerebral blood flow of 56 ± 10 ml/min/100 g. The bias in SvO2 of MOTIVE and BGS‐EPI to TRUST was +2 ± 4% and +1 ± 3%, respectively. The bias in cerebral blood flow of MOTIVE to Cartesian phase‐contrast reference was +1 ± 6 ml/min/100 g.Conclusions: The MOTIVE sequence is an advance over existing T2‐based oximetric methods. It does not require a control image and simultaneously measures SvO2 and flow velocity. The measurements agree well with TRUST and reference phase‐contrast sequences. This noninvasive technique enables CMRO2 quantification in under 20 s and is reproducible for in vivo applications.Keywords
Funding Information
- National Institutes of Health (F30 DK130510, P41 EB029460, T32 EB020087)
- National Center for Advancing Translational Sciences
- Institute for Translational Medicine and Therapeutics
This publication has 33 references indexed in Scilit:
- High temporal resolution in vivo blood oximetry via projection-basedT2measurementMagnetic Resonance in Medicine, 2012
- Blood oxygenation level‐dependent (BOLD)‐based techniques for the quantification of brain hemodynamic and metabolic properties – theoretical models and experimental approachesNMR in Biomedicine, 2012
- Test–retest reproducibility of a rapid method to measure brain oxygen metabolismMagnetic Resonance in Medicine, 2012
- Optimization of background suppression for arterial spin labeling perfusion imagingMagnetic Resonance Materials in Physics, Biology and Medicine, 2011
- QUantitative Imaging of eXtraction of oxygen and TIssue consumption (QUIXOTIC) using venular‐targeted velocity‐selective spin labelingMagnetic Resonance in Medicine, 2011
- Calibration and validation of TRUST MRI for the estimation of cerebral blood oxygenationMagnetic Resonance in Medicine, 2011
- MRI Estimation of Global Brain Oxygen Consumption RateJournal of Cerebral Blood Flow & Metabolism, 2010
- Noninvasive quantification of whole‐brain cerebral metabolic rate of oxygen (CMRO2) by MRIMagnetic Resonance in Medicine, 2009
- Quantitative evaluation of oxygenation in venous vessels using T2‐Relaxation‐Under‐Spin‐Tagging MRIMagnetic Resonance in Medicine, 2008
- Compensation for pulse imperfections in NMR spin-echo experimentsJournal of Magnetic Resonance (1969), 1981