Motion based adaptive calibration of pulse transit time measurements to arterial blood pressure for an autonomous, wearable blood pressure monitor
- 1 August 2008
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 2008 (1094687X) , 989-992
Abstract
This paper presents a novel adaptive algorithm for calibrating non-invasive pulse transit time (PTT) measurements to arterial blood pressure (BP). This new algorithm allows complete calibration of PTT to BP without the use of an oscillometric blood pressure cuff or external pressure sensor. Further, the algorithm can be used to continually update the identified parameters in the calibration equation while the patient is wearing the device. The technique utilizes natural patient motion to generate a known change in the transmural pressure (input) acting on the arteries monitored by our device to produce a measurable change in pulse transit time (output). The natural motion includes varying the height of the sensor relative to the heart to alter hydrostatic pressure at the measurement site and adjusting proximal joint posture to vary the external arterial pressure at the measurement site. This new algorithm is applied to a unique wearable sensor architecture that combines two in-line PPG sensors, one located at the ulnar artery of the wrist and one located at the digital artery of the little finger along with a multi-axis accelerometer for height measurement. Initial human subject tests results using the new algorithm and device will be presented.Keywords
This publication has 12 references indexed in Scilit:
- Adaptive hydrostatic blood pressure calibration: Development of a wearable, autonomous pulse wave velocity blood pressure monitor2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2007
- Theoretical Study on the Effect of Sensor Contact Force on Pulse Transit TimeIEEE Transactions on Biomedical Engineering, 2007
- A hydrostatic pressure approach to cuffless blood pressure monitoring2007
- Adaptive blood pressure estimation from wearable PPG sensors using peripheral artery pulse wave velocity measurements and multi-channel blind identification of local arterial dynamics2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2006
- The effect of applied sensor contact force on pulse transit timePhysiological Measurement, 2006
- Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressureJournal of Applied Physiology, 2006
- Prognostic Value of Ambulatory Blood-Pressure Recordings in Patients with Treated HypertensionNew England Journal of Medicine, 2003
- Clinical applications of arterial stiffness, Task Force III: recommendations for user proceduresAmerican Journal of Hypertension, 2002
- Home Blood Pressure Monitoring. Current Knowledge and Directions for Future ResearchBlood Pressure, 2001
- Measurements of Young's modulus of elasticity of the canine aorta with ultrasoundUltrasonic Imaging, 1979