RFID-Based Real-Time Navigation for Interventional Magnetic Resonance Imaging: Development and Evaluation of a Novel Tracking System
- 27 June 2017
- journal article
- research article
- Published by ASME International in Journal of Medical Devices
- Vol. 11 (3)
Abstract
The purpose of this study was to evaluate the suitability of a novel radio-frequency identification (RFID)-based tracking system for intraoperative magnetic resonance imaging (MRI). A RFID tracking system was modified to fulfill MRI-compatibility and tested according to ASTM and NEMA. The influence of the RFID tracking system on MRI was analyzed in a phantom study using a half-Fourier acquisition single-shot turbospin echo (HASTE) and true fast imaging with steady-state precession sequence (TrueFISP) sequence. The RFID antenna was gradually moved closer to the isocenter of the MR scanner from 90 to 210 cm to investigate the influence of the distance. Furthermore, the RF was gradually changed between 865 and 869 MHz for a distance of 90 cm, 150 cm, and 210 cm to the isocenter of the magnet to investigate the influence of the frequency. The specific spatial resolution was measured with and without a permanent line of sight (LOS). After the modification of the reader, no significant change of the signal-to-noise ratio (SNR) could be observed with increasing distance of the RFID tracking system to the isocenter of the MR scanner. Also, different radio frequencies of the RFID tracking system did not influence the SNR of the MR-images significantly. The specific spatial resolution deviated on average by 8.97 ± 7.33 mm with LOS and 11.23 ± 12.03 mm without LOS from the reference system. The RFID tracking system had no relevant influence on the MR-image quality. RFID tracking solved the LOS problem. However, the spatial accuracy of the RFID tracking system has to be improved for medical usage.Keywords
Funding Information
- Deutsche Forschungsgemeinschaft (TE 289/2-1)
This publication has 24 references indexed in Scilit:
- Evaluation of MRI issues for an access port with a Radiofrequency Identification (RFID) tagMagnetic Resonance Imaging, 2013
- Real‐time correction by optical tracking with integrated geometric distortion correction for reducing motion artifacts in functional MRIMagnetic Resonance in Medicine, 2012
- The impact of imaging speed of MR-guided punctures and interventions in static organs—A phantom studyEuropean Journal of Radiology, 2011
- Minimally Invasive Magnetic Resonance Imaging-Guided Free-Hand Aspiration of Symptomatic Nerve Route Compressing Lumbosacral Cysts Using a 1.0-Tesla Open Magnetic Resonance Imaging SystemCardioVascular and Interventional Radiology, 2011
- Safety and reliability of Radio Frequency Identification Devices in Magnetic Resonance Imaging and Computed TomographyPatient Safety in Surgery, 2010
- Modifikation einer kabellosen PC-Maus für die Steuerung der Bedieneroberfläche während der Intervention in der offenen Hochfeld-MRT bei 1,0 TRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2009
- Image-guided spinal injection procedures in open high-field MRI with vertical field orientation: feasibility and technical featuresEuropean Radiology, 2009
- Magnetic resonance imaging in the evaluation of pelvi‐ureteric junction obstruction: an all‐in‐one approachBJU International, 2007
- Interventional navigation for abdominal therapy based on simultaneous use of MRI and ultrasoundMedical & Biological Engineering & Computing, 2006
- Magnetic resonance imaging of freely moving objects: prospective real-time motion correction using an external optical motion tracking systemNeuroImage, 2006