Validation of the International Consensus Group Criteria for Slide Review Following Automated Complete Blood Count at Jimma Medical Center, Ethiopia
Open Access
- 1 March 2023
- journal article
- review article
- Published by Taylor & Francis in Journal of Blood Medicine
- Vol. ume 14, 213-220
Abstract
Background: The international consensus group suggested criteria for action following automated complete blood count and white blood cell differential analysis. These criteria were set based on data from laboratories of developed countries. It is highly important to validate the criteria in developing countries where infectious diseases are still rampant that can affect blood cell count and morphology. Thus, this study aimed to validate the consensus group criteria for slide review at Jimma Medical Center, Ethiopia from November 1, 2020 to February 30, 2021. Methods: The study comprised a total of 1685 patient samples from the daily laboratory workload of CBC analysis. The samples were collected in K2-EDTA tubes (Becton Dickinson) and analyzed using Coulter DxH 800 and Sysmex XT-1880 hematology analyzers. A slide review was done on two Wright-stained slides for each sample. All statistical analyses were performed using SPSS version 20 software. Results: There were 39.8% positive findings, the majority of which were related to red blood cells. The false negative and false positive rates for Sysmex and Coulter analyzer were 2.4% vs 4.8%; and 4.6% vs 4.7%, respectively. The false negative rate was unacceptably higher when we used physicians’ triggered slide review, which was 17.3% and 17.9% for Sysmex and Coulter analyzers, respectively. Conclusion: Generally, the consensus group rules are suitable to use in our setting. However, we might still need to modify the rules, particularly to reduce the review rates. It is also necessary to confirm the rules with case mixes proportionally derived from the source population.Keywords
This publication has 14 references indexed in Scilit:
- Comparing the performance of three panels rules of blood smear review criteria on an Italian multicenter evaluationInternational Journal of Laboratory Hematology, 2017
- Validation rules for blood smear revision after automated hematological testing using Mindray CAL-8000Journal of Clinical Laboratory Analysis, 2016
- Neutrophil left shift and white blood cell count as markers of bacterial infectionClinica Chimica Acta, 2016
- PERIPHERAL BLOOD FILM - A REVIEW2014
- Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories?Revista Brasileira de Ginecologia e Obstetrícia, 2014
- Validation and Optimization of Criteria for Manual Smear Review Following Automated Blood Cell Analysis in a Large University HospitalArchives of Pathology & Laboratory Medicine, 2013
- Comparison of haematological parameters determined by the Sysmex KX - 2IN automated haematology analyzer and the manual countsBMC Clinical Pathology, 2010
- Laboratory productivity and the rate of manual peripheral blood smear review: a College of American Pathologists Q-Probes study of 95,141 complete blood count determinations performed in 263 institutions.Archives of Pathology & Laboratory Medicine, 2006
- Diagnosis from the Blood SmearNew England Journal of Medicine, 2005
- The International Consensus Group for Hematology Review: Suggested Criteria for Action Following Automated CBC and WBC Differential AnalysisLaboratory Hematology, 2005