Electrochemical characterization of a carbon sensor for ascorbic acid detection

Authors:

Abstract
A commercial screen-printed carbon electrode was employed as a low-cost platform for the detection of ascorbic acid (AA). The surface of the working electrode and solutions of AA were characterized by Raman and UV–Vis analysis. The surface of the working and counter electrodes is composed of carbon, while the surface of the reference electrode is made of polyester. The surface of the electrode became covered with material when in contact with AA, and EDS showed a decrease in the content of carbon in the electrode area and an increase in the content of oxygen and of elements that are a part of the PBS buffer solution. The UV–Vis spectra of AA showed an absorption band at wavelengths of 265–270 nm, which increased in intensity with increase in concentration. The electrochemical behavior of AA was found to involve an irreversible oxidation and a broad anodic peak with no cathodic peak. The current of the anodic peak varied linearly with the square root of the scan rate, indicating that the mass transport process is controlled by a diffusion mechanism. Chronoamperometry provided stepwise current responses that varied linearly with the concentration of AA. A calibration plot was obtained with R 2 = 0.997 for the concentration range of 10–100 µM. Chronoamperometry was found to be the most convenient method for the detection of AA using the commercial screen-printed carbon electrode.

This publication has 21 references indexed in Scilit: