Abstract
We discuss the design and performance of a liquid nitrogen-cooled optical microscopy system that also allows for simultaneous transport measurements. With a cold-finger based design, the system operates at temperature between 100 K to 400 K with controlled ramp rate (0.1 K/min to 42 K/min) and thermal stability (±5 mK) at any given temperature. The cryostat forms a part of a modular and versatile experimental set up, particularly suitable for low temperature bright-field reflection microscopy under Köhler illumination. With careful selection of optical components, an active focus correction system to account for the thermal expansion, along with effective vibration isolation, a sub-micron spatial resolution has been achieved. Finally, we demonstrate optical imaging of the electric breakdown in vanadium dioxide using this setup.
Funding Information
  • Indian Institute of Science Education and Research Kolkata
  • ANRF, Department of Science and Technology, Government of India (CRG/2022/008662)

This publication has 80 references indexed in Scilit: