Biodegradable, fully threaded magnesium alloy headless compression screw for small-bone fixation: Mechanical and biological safety assessments and in vivo study
Headless compression (HC) screws used in the treatment of bone fractures have clinical issues related to screw material and design, including loosening and adverse effects from long-term implantation. Consequently, there is a strong need for biodegradable HC screws in clinical sites. In this study, we conducted safety assessments and in vivo tests on a newly developed, fully threaded, biodegradable magnesium (Mg) alloy HC screw to evaluate its feasibility for clinical use. Mechanical and biological safety assessment results indicated that the developed Mg alloy screw demonstrated insertability and bone fixation performance, as well as excellent biocompatibility, comparable to or superior to that of conventional approved HC screws. In vivo studies using a rabbit model over short- and long-term periods revealed that the developed Mg alloy screw was almost completely degraded and resorbed by 10 months post-implantation. During the degradation process, gas cavities formed around the implanted screw and subsequently decreased, accompanied by bone formation and remodeling. Furthermore, the effect of screw degradation products on target organs was found to be minimal. These findings indicated that the developed full-thread Mg alloy HC screw identifies no safety concerns for clinical use.