A user-friendly visual brain-computer interface based on high-frequency steady-state visual evoked fields recorded by OPM-MEG
- 30 May 2024
- journal article
- research article
- Published by IOP Publishing in Journal of Neural Engineering
- Vol. 21 (3) , 036024
Abstract
Objective. Magnetoencephalography (MEG) shares a comparable time resolution with electroencephalography. However, MEG excels in spatial resolution, enabling it to capture even the subtlest and weakest brain signals for brain-computer interfaces (BCIs). Leveraging MEG’s capabilities, specifically with optically pumped magnetometers (OPM-MEG), proves to be a promising avenue for advancing MEG-BCIs, owing to its exceptional sensitivity and portability. This study harnesses the power of high-frequency steady-state visual evoked fields (SSVEFs) to build an MEG-BCI system that is flickering-imperceptible, user-friendly, and highly accurate. Approach. We have constructed a nine-command BCI that operates on high-frequency SSVEF (58–62 Hz with a 0.5 Hz interval) stimulation. We achieved this by placing the light source inside and outside the magnetic shielding room, ensuring compliance with non-magnetic and visual stimulus presentation requirements. Five participants took part in offline experiments, during which we collected six-channel multi-dimensional MEG signals along both the vertical (Z-axis) and tangential (Y-axis) components. Our approach leveraged the ensemble task-related component analysis algorithm for SSVEF identification and system performance evaluation. Main Results. The offline average accuracy of our proposed system reached an impressive 92.98% when considering multi-dimensional conjoint analysis using data from both the Z and Y axes. Our method achieved a theoretical average information transfer rate (ITR) of 58.36 bits min−1 with a data length of 0.7 s, and the highest individual ITR reached an impressive 63.75 bits min−1. Significance. This study marks the first exploration of high-frequency SSVEF-BCI based on OPM-MEG. These results underscore the potential and feasibility of MEG in detecting subtle brain signals, offering both theoretical insights and practical value in advancing the development and application of MEG in BCI systems.Keywords
Funding Information
- National Natural Science Foundation of China (61976152)
- China Postdoctoral Science Foundation (2021M702445)
This publication has 38 references indexed in Scilit:
- A hybrid steady-state visual evoked response-based brain-computer interface with MEG and EEGExpert Systems with Applications, 2023
- Optically Pumped Magnetometers for Practical MEG-Based Brain-Computer Interfacing2021
- Rapid invisible frequency tagging reveals nonlinear integration of auditory and visual informationHuman Brain Mapping, 2020
- Wearable neuroimaging: Combining and contrasting magnetoencephalography and electroencephalographyNeuroImage, 2019
- Optically pumped magnetometers: From quantum origins to multi-channel magnetoencephalographyNeuroImage, 2019
- A bi-planar coil system for nulling background magnetic fields in scalp mounted magnetoencephalographyNeuroImage, 2018
- IFCN-endorsed practical guidelines for clinical magnetoencephalography (MEG)Clinical Neurophysiology, 2018
- A new generation of magnetoencephalography: Room temperature measurements using optically-pumped magnetometersNeuroImage, 2017
- Use of high-frequency visual stimuli above the critical flicker frequency in a SSVEP-based BMIClinical Neurophysiology, 2015
- Human EEG responses to 1?100 Hz flicker: resonance phenomena in visual cortex and their potential correlation to cognitive phenomenaExperimental Brain Research, 2001