Assessment of deformation of human red blood cells in flow cytometry: measurement and simulation of bimodal forward scatter distributions
Authors:
- 8 August 2019
- journal article
- research article
- Published by Optica Publishing Group in Biomedical Optics Express
- Vol. 10 (9) , 4531-4550
Abstract
Light scattering by single cells is widely applied for flow cytometric differentiation of cells. However, even for human red blood cells (RBCs), which can be modeled as homogeneous dielectric particles, the potential of light scattering is not yet fully exploited. We developed a dedicated flow cytometer to simultaneously observe the forward scattering cross section (FSC) of RBCs for orthogonal laser beams with incident wave vectors k → 1 and k → 2 . At a wavelength λ = 632.8 nm , bimodal distributions are observed in two-dimensional dot plots of FSC( k → 1 ) vs. FSC( k → 2 ), which result from the RBCs' random orientation around the direction of flow, as well as from the distributions of their size and their optical properties. Typically, signals of 7.5 × 10 4 RBCs were analyzed. We actively oriented the cells in the cytometer to prove that orientation is the main cause of bimodality. In addition, we studied the wavelength dependence of FSC( k → 1 ) using λ = 413.1 nm , 457.9 nm , 488 nm and 632.8 nm, covering both weak and strong light absorption by the RBCs. Simulations of the light scattering by single RBCs were performed using the discrete dipole approximation (DDA) for a range of sizes, orientations and optical properties to obtain FSC distributions from RBC ensembles. Using the axisymmetric biconcave equilibrium shape of native RBCs, the experimentally observed distributions cannot be reproduced. If, however, an elongated shape model is employed that accounts for the stretching of the cell by hydrodynamic forces in the cytometer, the features of the strongly bimodal measured frequency distributions are reproduced by the simulation. Elongation ratios significantly greater than 1 in the range of 1.5 to 2.5 yield the best agreement between experiments and simulated data.Keywords
This publication has 42 references indexed in Scilit:
- Flow Cytometric Assessment of Erythrocyte Shape through Analysis of FSC Histograms: Use of Kurtosis and Implications for Longitudinal EvaluationPLOS ONE, 2013
- The discrete-dipole-approximation code ADDA: Capabilities and known limitationsJournal of Quantitative Spectroscopy and Radiative Transfer, 2011
- Red blood cell deformation in microconfined flowSoft Matter, 2009
- Extensional flow-based assessment of red blood cell deformability using hyperbolic converging microchannelBiomedical Microdevices, 2009
- Effect of plasmodial RESA protein on deformability of human red blood cells harboring Plasmodium falciparumProceedings of the National Academy of Sciences, 2007
- Assessment of erythrocyte shape by flow cytometry techniquesJournal of Clinical Pathology, 2006
- Deformability‐based flow cytometryCytometry Part A, 2004
- Plasmodium falciparum Maturation Abolishes Physiologic Red Cell DeformabilityScience, 1984
- Isovolumetric sphering of erythrocytes for more accurate and precise cell volume measurement by flow cytometryCytometry, 1983
- Electromagnetic Field of a Gaussian Beam with an Elliptical Cross SectionJournal of the Optical Society of America, 1972