Abstract
Purpose: To prospectively evaluate the recently introduced international standard method for measurement of the detective quantum efficiency (DQE) of digital radiography systems, in comparison with representative prior methods. Materials and Methods: A recently introduced international standard method (International Electrotechnical Commission [IEC] 62220-1, 2003) for DQE measurement and two previously described DQE evaluation methods were considered. In addition to an overall comparison, evaluations of the following method factors were performed: beam quality, beam-limiting devices (apertures or collimators), noise power spectrum (NPS) analysis algorithms and parameters (area, region of interest size, background detrending), and modulation transfer function (MTF) test devices and methods. Results: Overall, at low to middle frequencies, the IEC method yielded DQE estimates that were 3.3% and 6.5% lower than the values yielded by the two previous methods. Averaged over the frequency range of 1.5–2.5 mm−1, ...